XDCQ-5 type hydrocarbon element analyzer
The rapid automatic hydrogen analyzer is mainly used to determine the content of hydrogen element in coal and other solid materials. The entire testin
Product details
Technical specifications and parameters of XDCQ-5 rapid automatic hydrogen analyzer
Measurement range: Hydrogen 0-20%, Carbon 1-100%
Heating rate: 25 ℃/min, 30 minutes or rising to 800 ℃
Measurement time: approximately 10-15 minutes
Heating control: 25K/minute, rise to 600 ℃ in 30 minutes; Linearity:+ 0.1%
Allowable error: Hydrogen: ≤ 0.15% (same laboratory)
Carbon: ≤ 0.50%
Power supply: AC220V ± 10%, 50Hz
Temperature control: 800 ℃± 10 ℃, 300 ℃± 10 ℃
Linearity: ± 0.1%
Sample: Particle size<0.2mm
Weight: Approximately 65mg
Accuracy and precision: comply with the requirements of the national standard GB/T476-2001.