Y09-310 High Flow Laser Dust Particle Counter
Y09-310 laser dust particle counter: (high flow rate) used to measure the size and number of dust particles per unit volume of air in clean environmen
Product details
Y09-310typeLarge trafficlaser particle counter
Y09-310Laser dust particle counter
Y09-310Laser dust particle counter:(High flow rate) is used to measure the size and number of dust particles per unit volume of air in clean environments, and can directly detect clean environments with cleanliness levels ranging from 300000 to 100. This instrument adopts a semiconductor laser light source, a large LCD screen display, high detection accuracy, simple and clear functional operation, microprocessor control, and can store and print measurement data. It is very convenient for testing in a clean environment. In addition, the data collected by the dust particle counter can be uploaded to the computer through dedicated communication software, and can be edited, saved, and printed in Excel in Office. This product is widely used in sectors such as electronics, optics, chemistry, food, cosmetics, healthcare, biological products, aerospace, etc.
Main technical parameters
External dimensions
300 x 360 x 126mm3(Width x Depth x Height)
quality
6.5kg
Maximum power consumption
60W
Power supply
AC power supply 220V ± 10%, 50Hz
Particle size channel
0.3, 0.5, 1, 3, 5, 10 (μ m), six particle sizes can be displayed simultaneously, and particle concentration (unit: particle/cubic meter) can also be directly displayed and printed
Sampling traffic
28.3L/min
sampling period
1 second to 24 hours continuously adjustable
Delay Time
0-60S
Continuous sampling frequency
0-99 times
Interval time
0-99 points
clean-down capability
≤10min
printer
Embedded thermal printer
Alarm (cleanliness level)
10. Level 100, 1000
Confidence calculation function
UCL calculation function
Environmental conditions for use
Temperature: 10 ℃ -30 ℃
Humidity: 20% -75%
Atmospheric pressure: 86kPa-16kPa
Maximum allowable sampling concentration
35000 particles/L (particle size ≤ 0.5 μ m), the sampled air must not contain corrosive gases such as acid and alkali
Extend working hours
8 hours
Light source and lifespan
Semiconductor laser with a lifespan of over 30000 hours
Communication software
Software installation CD, data communication cable
data storage
1000 sets