YAMATO Organic Solvent Recovery Unit GAS410
It is connected with the spray dryer to form a nitrogen closed circulation mode with high safety.
When organic solvents are used in spray dryers ADL311S, GB210-A and DL410, organic solvent recovery devices are used to prevent external discharge.
The treatment of flammable and toxic solvents can be achieved by using a nitrogen sealed circulation method and a combination of a refrigeration unit and a condenser for solvent recovery.
It can dry easily oxidizable substances.
Due to its ability to dry at low temperatures, it can also dry substances that are prone to thermal deformation.
Under strict safety measures, spray drying is carried out to recover products and solvents.
Specifications:
model |
GAS410 |
|
Solvent recovery method |
Condenser+freezer |
|
performance |
Circulating gas |
Nitrogen (connected to ADL311S-A, GB210-A, DL410 in a closed loop) |
Circulating air volume |
0.12~0.65m3/min |
|
constitute |
Air compressor (for spray) |
Linear reciprocating air compressor |
Circulating blower |
Roots blower |
|
Solvent recovery container |
2L flask |
|
freezer |
Air cooled fully enclosed, 400W R404A |
|
Solvent recovery structure |
Condenser cooling structure |
|
filter |
Filter cartridge type cooling structure |
|
operation panel |
Condenser temperature display monitor |
|
Filter differential pressure gauge (filter mesh blockage monitoring) | ||
Oxygen concentration monitor | ||
Blower air volume adjustment knob | ||
Oxygen sensor |
Fuel cell sensor |
|
pump |
Oxygen sensor cycling |
|
Safety device |
Oxygen concentration meter, combustible gas alarm, overcurrent leakage protection switch, N2 forced introduction (when nozzle is removed) |
|
Specifications |
External dimensions |
710×D950×H1450mm |
weight |
About 130kg |
|
Rated current of power supply (50/60Hz) |
AC200~240V 5A |
|
Necessary N2 quantity |
When set at 0.2MPa, 30L/h |
|
Accessories |
Connect one set of related components, 2L flask |