Small desktop vacuum drying oven designed for various purposes of vacuum drying
Through the auxiliary menu key, operations such as over lift preventer, deviation correction, and key lock setting can be achieved.
● Equipped with self diagnostic circuit (temperature sensing abnormality, heater disconnection, automatic over rise prevention, SSR short circuit), over rise protector, leakage protection switch to prevent overcurrent, button lock and other safety functions.
For safety reasons, a protective panel made of resin is installed on the observation window.
■Specifications
model |
DP23C |
DP33C |
mode |
Decompression and wall heating |
|
Temperature range for use |
40~200℃ |
|
Use pressure range |
101~0.1kPa(760~1Torr) |
|
Maximum temperature arrival time |
About 60 minutes |
About 90 minutes |
Temperature regulation accuracy |
±1.5℃(at240℃) |
|
Heating method |
Direct heating of the pressure groove wall |
|
Heater power |
0.68KW |
1.05KW |
timer |
1 minute to 99 hours and 59 minutes, and 100 to 999 hours and 50 minutes (with timed waiting function) |
|
Safety device |
Self diagnostic circuit (temperature sensing abnormality, heater disconnection, automatic over rise prevention, SSR short circuit), over rise protector, overcurrent leakage protection switch, button lock function |
|
Internal dimensions (width x depth x height mm) |
200×250×200 |
300×300×300 |
Dimensions (width x depth x height mm) |
400×410×672 |
510×460×774 |
Internal volume |
10L |
27L |
Number of layers/spacing of shed boards |
3 layers (fixed)/63mm |
4 layers (fixed)/71mm |
weight |
About 43kg |
About 69kg |
Accessories |
Stainless steel stamping mesh plate, 2 pieces |
|
Product selection |
Shed board, vacuum pump, N2 inlet device, recorder, combination warning light (standby/running/fault), external communication function (RS485), temperature output terminal (4-20mA), external alarm output terminal, time arrival output terminal |