Specification parameters
Temperature range:- 40℃~150℃
Minimum temperature: -55 ℃
Humidity range: 20% to 98% RH
Heating time:- 40 ℃~+100 ℃ within 45 minutes (when unloaded)
Cooling time:+ 20 ℃~-40 ℃ within 45 minutes (unloaded)
Control stability: temperature ± 1.0 ℃, humidity ± 1.0% RH
Control uniformity: temperature ± 2.0 ℃, humidity ± 3.0% RH
Power Supply: AC、 Single phase, 50/60HZ, 220V, 3.5KW
structure
Size:
a. Internal dimensions: W40 × H50 × D50cm
b. External dimensions: W65 × H140 × D115cm
c. Space requirement: W140 × H200 × D137cm
texture of material:
a. Inner box material: Mirror stainless steel plate (SUS # 304)
b. Outer box material: First grade steel plate electrostatic spraying
c. Base material: National standard angle iron
d. Insulation material: Hard Polurethane foam+glass wool
Supply air circulation system:
a. 1 60W motor
b. Extended axis
c. Multi wing wind turbine
d. Special adjustable hundred page air supply circulation system
Box door: single door, left door, handle on the right-hand side
Window: Three layers of vacuum
Door buckle: embedded non reactive flat handle
Rear button: SUS#304
Window light: 5W
Test hole: 1 50mm hole on the left side of the body, with 1 stainless steel hole cover and 1 silicone plug attached
Refrigeration system
Compressor: French Taikang fully enclosed compressor (2 sets of stacked 375W)
Refrigerant: Environmentally friendly refrigerant R-404A
Condenser: air-cooled condenser
Evaporator: finned evaporator
Heating tube: upturned 304 stainless steel, 1KW
Other original components: Dry filter (American ALCO), solenoid valve (Japanese Egyo)
Refrigeration components such as oil separators (American ALCO) are all imported
Control system
Controller: Adopting a 7-inch color touch screen controller
(1) Controller specifications:
Accuracy: Temperature ± 0.1 ℃+1 digit; Relative humidity ± 0.1%+1 digit
Equipped with upper and lower limit standby and alarm functions
6 groups of P I. D control parameter setting, P.I.D automatic calculation
Equipped with power-off program memory, it automatically starts and continues to execute program functions after power restoration
Equipped with reservation activation function; Date and time adjustment function
It has communication interface control function (software can be distributed)
(2) Screen display function:
Chinese and English button operation
It can display the current program execution number, segment number, remaining time, and number of cycles
Accumulated operation time function
The temperature and humidity program settings are displayed in graphical curves, with real-time display of program curve execution function
The fault prompt screen is displayed
Key locking function
(3) Program capacity and control functions:
Program capacity: Up to 12 program groups, 120 program segments, each segment can be set for a time of 0-99Hour59Min
Program Loop: Each program group can loop 999 times, and partial loops and program links can also be performed
Program editing: It has functions such as editing, clearing, and inserting, and can input 4 segments of temperature and time per page
Machine features
*-At temperatures above 20 degrees, only one 375W compressor is used in conjunction with a 1KW heating tube for incremental heating, saving energy and water; If single-stage refrigeration is used, the compressor is 1.5 horsepower, which is 1125W, and works with a 2.2KW heating tube. It has a large cooling capacity and strong dehumidification function, so it consumes only 750W of electricity and does not save water. When dual compressors are started, it is only 750W.
*The sink is made of special materials with low thermal conductivity and no water droplets on the surface (if metal materials are used, water droplets may form on the surface, which can easily wet the sample and affect the testing effect). It is also equipped with a flow guide device. If the internal pressure or water level of the sink is too high, it will flow into the humidification bucket through the flow guide hole.
*Easily achieve a low temperature of -50 degrees and a maximum temperature of -55 degrees. When single-stage refrigeration reaches -37 degrees, it is difficult to further reduce to -40 degrees due to the characteristics of R404A refrigerant, which can only be achieved through temperature compensation. Third party inspection of -40 degrees may reveal an error of 2 to 3 degrees.
*The machine is exquisite, with a large internal space and small external dimensions, with a width of only about 60CM. The single door laboratory can be easily accessed.
*There is a pressure relief hole attached to the top.
Standard configuration
Two pieces of stainless steel SUS # 304 grid sample placement rack, with each layer of the sample weighing no more than 10 kilograms
One spare special wet cloth