As one of the existing functions of the IOL-IMAGER imaging measurement instrument, the brief introduction here is the same as the IOL-IMAGER imaging measurement instrument * and is only for online promotion purposes. Please forgive us!
- The measurement principle of the artificial lens optical eccentricity tester is to place the artificial lens in a fixture with a prescription diameter (10 or 11mm) and measure the offset distance between the geometric center of its optical body and the center of the fixture hole seat.
- The optical eccentricity measurement function of artificial intraocular lens is already included in IM-3 and does not need to be purchased separately.
Artificial intraocular lens imaging measurement instrument IM-3Measurable and Observable:
- Measurable physical properties specified in ISO11979-3 or YY0290.3: axial displacement, optical eccentricity, optical tilt angle (measurable in four directions), contact angle
- Measurable: Lens diameter, outer contour, loop diameter, length, and angle, etc
- Observable: Lens surface quality, defects, knife marks, edge burrs, etc
Artificial intraocular lens imaging measurement instrument IM-3Product advantages:
- One instrument can perform both frontal and lateral measurements.
- XYZ three-axis can move quickly, Z-axis can be fine tuned.
- Provide four independently controlled directional lighting modes: front lighting, back lighting, side lighting, and diagonal lighting.
- The entire measurement process is carried out in a constant temperature water bath, simulating the human eye environment.
- High definition cameras and optical paths provide high-quality images.
- Powerful software provides various measurement and export functions.
Host specification parameters:
- Size: Width 330mm, Depth 230mm, Height 515mm
- Weight: Approximately 11Kg
- Output: USB3.0-B
- Mobile range:
X-axis (left-right movement): 60mm
Y-axis (forward and backward movement): 50mm
Z-axis (up and down focus): 55mm+10mm fine adjustment
Temperature control/Specification parameters of light source control box:
- model: TC-35-IM
- Size: Width 150mm, Depth 235mm, Height 120mm
- Weight: Approximately 1.8kg
- Input: 220VAC 2A
- Output (large): 5V-1000mA, 0~12V-800mA, 24V-2500mA
- Positive lighting: one level adjustable
- Side lighting: adjustable in two levels
- Backlight: Two adjustable levels
- Slant lighting: one fixed level
- Temperature display resolution: 0.1 ℃
Optical parameters:
- Sensor size: 1/2.5 CMOS (5.76 * 4.29mm)
- Resolution: 5 million pixels (2592 * 1944)
- Pixel size: 2.2um * 2.2um
- Shutter: Electronic Rolling Shutter
- Sensitivity: 1.76V/lux-sec 550nm
- Object distance: 110mm
- Field of view: approximately 19.2mm x 14.4mm
- Optical magnification: 0.3X
- Display magnification: Depending on the monitor (approximately 38 times for a 1:1 22 inch display and 109 times for enlargement)
- Positive lighting: white coaxial downward illumination, adjustable brightness
- Side lighting: white LED soft light board, adjustable angle from 0 to 90 degrees, brightness adjustable and switchable
- Back lighting: white LED soft light board, adjustable brightness and switchable
- Slant lighting: white LED spotlight, adjustable angle at will
Measurement slots and fixtures:
- Heating method: Ceramic heating
- Heating power: up to 60W
- Medium used: water or salt solution
- Fixture categories: standard fixtures, circular quick fixtures, bidirectional fixtures, square quick fixtures, a total of 7 in four types
- Fixture clamping: stroke of 13mm, resolution of 0.001mm, full cycle repeatability accuracy better than 0.02mm
Measurement software parameters:
- Measurement functions: distance, angle, trajectory, area, radius, perimeter
- Image processing: brightness, contrast, sharpness, color, negative film, rotation, flipping, straightening
- Image scaling: 1%~800%
- The measurement results can be exported as an Excel file
- Repeated measurements can be analyzed and statistically analyzed
- Default measurement resolution: 0.001mm