Automatic refractometer RA-620/RA-600 for petroleum products
The fully automatic refractometer RA-620/RA-600 for petroleum products meets the standard:
GB/T 6488-2008 Determination of refractive index of liquid chemical products (20 ° C).
GB/T 614-2006 Chemical reagents General method for determination of refractive index.
GB/T 5527-2010 Determination of refractive index of animal and vegetable oils and fats.
GB/T 13531.7-2018 General inspection methods for cosmetics - Determination of refractive index.
SH/T 0724-2002 Determination of refractive index and refractive dispersion of liquid hydrocarbons.
SH/T 0205-1992 Determination of refractive index and specific dispersion of electrical insulating liquids.
SH/T 0729-2004 Calculation method for carbon distribution and structural group composition of petroleum fractions (n-d-M method).
NB/SH/T 0838-2010 Determination of polycyclic aromatic hydrocarbons (PCA) content in unused lubricating base oils and asphaltene free petroleum fractions by dimethyl sulfoxide extraction refractive index method.
SN/T 0041-2011 Test method for PONA value of exported naphtha Part 4 Determination of cycloalkanes in saturated hydrocarbon fractions by refractive index method.
MH/T 6114-2017 Determination of refractive index of de icing and anti icing fluids.
ASTM D1218 - 12(2016) Standard Test Method for Refractive Index and Refractive Dispersion of Hydrocarbon Liquids( Test method for refractive index and refractive dispersion of liquid hydrocarbons.
ASTM D1747-09 (2019) Standard Test Method for Refractive Index of Viscous Materials.
ASTM D1807 - 00(2005)e1 Standard Test Methods for Refractive Index and Specific Optical Dispersion of Electrical Insulating Liquids( Test method for refractive index and specific optical divergence of electrically insulating liquids.
ASTM D5006 - 11(2016) Standard Test Method for Measurement of Fuel System Icing Inhibitors (Ether Type) in Aviation Fuels( Test method for measuring fuel system icing inhibitors (ether type) in aviation fuels.
The main features of the fully automatic refractometer RA-620/RA-600 for petroleum products are:
1. Refractometer with touch screen LCD display, displaying refractive index, soluble solids, and concentration.
2. Built in fully automatic temperature control, total reflection measurement, easy operation, and fast measurement speed.
3. Using LED light source does not require replacing the bulb, and the measuring prism is sapphire wear-resistant and corrosion-resistant.
4. Digital refractometer, compliant with GLP standards, automatically stores results, inspection and calibration records.
5. The refractive index and concentration comparison table or calculation formula can be inputted to automatically convert the concentration value of the sample.
7. Standard sample non-volatile cover plate can prevent the impact of volatile samples during testing.
8. It can be connected to an external digital densitometer and various automatic samplers to simultaneously measure density and refractive index.
9. Chinese interface, simple and intuitive operation, high-precision measurement results, and ultra-high cost-effectiveness.
Technical parameters of the fully automatic refractometer RA-620/RA-600 for petroleum products:
form |
Fully automatic refractometer |
|||
Instrument model |
RA-620 |
RA-600 |
||
measuring method |
Detection of critical angle of light refraction |
|||
light source |
LED sodium spectrum D-line (589.3nm) |
|||
Measurement project |
Refractive index, sugar content, concentration |
|||
measuring range |
Refractive index (nD) |
1.32000~1.58000 |
1.3200~1.7000 |
|
Sugar content (Brix) |
0.00~100.00% |
|||
Accuracy |
Refractive index (nD) |
±0.00002 |
±0.0001 |
|
Sugar content (Brix) |
±0.014% |
±0.1% |
||
Repeatability |
Refractive index (nD) |
±0.00001 |
±0.0001 |
|
Sugar content (Brix) |
±0.007%(<5%), ±0.01%(≧5%) |
±0.1% |
||
Resolution |
Refractive index (nD) |
0.00001 |
0.0001 |
|
Sugar content (Brix) |
0.001%(<5%), 0.01%(≧5%) |
0.1% |
||
temperature control |
5-75 ° C, built-in Peltier temperature control (RA-600 can choose 5-100 ° C) |
|||
temperature resolution |
0.01°C |
0.1°C |
||
Sample size |
About 0.2mL |
|||
Sample contact material |
Prism |
sapphire |
||
Sample stand |
SUS316 stainless steel |
|||
display |
4.7-inch color LCD touch screen |
|||
Operation method |
Touch screen with touch pen |
|||
Security |
Password protection |
|||
Data storage |
measuring method |
100 methods |
||
measurement result |
300 sets of results |
|||
Correction record |
20 groups |
|||
Inspection records |
20 groups |
|||
External storage |
USB drive |
|||
Automatic temperature correction |
5~75°C Input the refractive index to a thermometer or polynomial, and automatically convert it to the refractive index at the desired temperature |
|||
Automatic concentration conversion |
100 sets Input refractive index to concentration table or polynomial, and automatically convert refractive index to concentration |
|||
Interface |
LAN: computer USB: USB flash drive, keyboard, barcode printer, printer RS-232C: Printer, automatic sample cleaning unit, multi sample automatic sampler |
|||
environment condition |
1) Temperature: 5~35°C 2) Humidity: Below 85% RH |
|||
Power Supply |
AC100~240V, 50Hz/60Hz (AC adapter), can be connected to an external rechargeable battery |
|||
power |
About 20 watts |
|||
size |
192(W) × 281(D) × 166(H)mm |
|||
weight |
About 5 kilograms |