Shanghai Huanao Technology Co., Ltd
Home>Products>BINDER ED23 drying oven
BINDER ED23 drying oven
BINDER ED23 drying oven
Product details

BINDER ED23 drying oven

1、 Product Introduction:

1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃

2 APT.line ™ Preheating chamber technology

3. Free convection

4. Adjustable exhaust valve

5. Controller with time function

6. 2 chrome plated sockets

7. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm

2、 Technical parameters:

Product model ED023-230V1 ED023-230V-I1 ED023UL-120V1

ED023UL-120V-I1

Choose a model standard Equipped with RS 422 standard Equipped with RS 422
Item code 9010-0190 9010-0191 9010-0192 9010-0193
measure
Internal volume [L] 20 20 20 20
Net weight of equipment (unloaded) [kg] 27 27 27 27
Maximum total load [kg] 25 25 25 25
Maximum load per partition [kg] 12 12 12 12
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 435 435 435 435
Net weight at height [mm] 495 495 495 495
Depth net weight [mm] 520 520 520 520
Wall distance from rear [mm] 100 100 100 100
Wall to side distance [mm] 100 100 100 100
Internal dimensions
Width [mm] 222 222 222 222
Height [mm] 330 330 330 330
Depth [mm] 300 300 300 300
The number of doors
Outer door 1 1 1 1
Data related to the environment
Energy consumption at 150 ℃ [Wh/h] 148 148 148 148
firmware
Number of partitions (hours/maximum) 2/4 2/4 2/4 2/4
Temperature performance data
The temperature range exceeds the indoor temperature by 5 ℃ to [℃] 300 300 300 300
Temperature deviation is 150 ℃ [± K] 2.8 2.8 2.8 2.8
Temperature fluctuation at 150 ℃ [± K] 0.5 0.5 0.5 0.5
Heating time to 150 ℃ [min] 28 28 28 28
The recovery time after opening the door in 30 seconds is 150 ℃ [min] 28 28 28 28
Ventilation data
Air exchange rate at 150 ℃ [x/h] 13 13 13 13
Electrical data
Rated voltage [V] 230 230 120 120
Power frequency [Hz] 50/60 50/60 60 60
Rated power [kW] 0.8 0.8 0.8 0.8
Equipment safety device [A] 10 10 12.5 12.5
Phase (rated voltage) 1~ 1~ 1~ 1~

1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.