Shanghai Huanao Technology Co., Ltd
Home>Products>BINDER MKFT720 High and Low Temperature Alternating Climate Chamber
BINDER MKFT720 High and Low Temperature Alternating Climate Chamber
BINDER MKFT720 High and Low Temperature Alternating Climate Chamber
Product details

BINDER MKFT720 High and Low Temperature Alternating Climate Chamber

1、 Product Introduction:

1. Temperature range: -70 ℃ to 180 ℃

2. Humidity range: 10% relative humidity to 98% relative humidity

3. Integrated water storage tank 20 L

4. Four zero voltage switch contacts that can be activated through the MCS controller

5 APT.lineTMPreheating chamber technology

6. Programmable condensation protection for samples

7. Heated observation window with LED internal lighting

8. Using capacitive humidity sensors and steam humidification for humidity regulation

9. BINDER communication software APT-COMTM 3 Basic version

10. Fault diagnosis system using optical and sound alarms

11. Controller using interval programming

12. 2 access holes with silicone plugs, 80mm, left and right

13. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm

14. 4 casters, 2 with brakes

15. Computer interface: Ethernet

16. 230 V socket on the right side operating component

17. Display through LCD color monitor

18. Adjustable slope function function function

19. Integrated continuous automatic recorder

20. Real time clock

21. Door heating

22. Alarm message when fresh water tank is short of water

23. A bracket made of stainless steel

24. Hose protection joint, including water supply and drainage pipes with a height not exceeding 1m

25. Inner chamber made of stainless steel

26. PT100 temperature sensor

27. Heat insulation through polyurethane foam

28. Fluorine free refrigerants R-404A and R-23

29. Cooling through cascaded compression cooling machines

30. Finned evaporator

2、 Technical parameters:

Product model MKFT720-400V1 MKFT720-400V-C1
Choose a model standard Equipped with transformer and frequency converter
Item code 9020-0288 9020-0301
measure
Internal volume [L] 734 734
Net weight of equipment (unloaded) [kg] 635 635
Maximum total load [kg] 160 160
Maximum load per partition [kg] 40 40
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 1580 1580
Net weight at height [mm] 2005 2005
Depth net weight [mm] 1140 1140
Wall to side distance [mm] 300 300
Wall distance from rear [mm] 335 335
Observation window width [mm] 508 508
Observation window height [mm] 300 300
Internal dimensions
Width [mm] 1200 1200
Height [mm] 1020 1020
Depth [mm] 600 600
The number of doors
Outer door 1 1
Data related to the environment
Sound pressure level [dB (A)] 69 69
firmware
Number of partitions (hours/maximum) 1/11 1/11
Temperature performance data
Temperature range [℃] -70…180 -70…180
The temperature deviation depends on the set value [± K] 0.3…2 0.3…2
The temperature fluctuation depends on the set value [± K] 0.1…0.5 0.1…0.5
The average heating rate is based on IEC 60068-3-5 [K/min] 4.8 4.8
The average cooling rate is based on IEC 60068-3-5 [K/min] 4 4
The maximum thermal compensation is at 25 ℃ and 90% relative humidity [W] 800 800
The maximum thermal compensation is at 25 ℃ [W] 5000 5000
Climate performance data
Temperature range [℃] 10…95 10…95
The temperature fluctuation depends on the set value [± K] 0.1…1 0.1…1
Humidity range [% relative humidity] 10…98 10…98
The humidity fluctuation depends on the set value ≤ 2.5 ± relative humidity ≤ 2.5 ± relative humidity
Condensation temperature range of condensate water [℃] 5…94 5…94
Electrical data
Rated voltage [V] 400 480
Power frequency [Hz] 50 60
Rated power [kW] 13 13
Equipment safety device [A] 32 32
Phase (rated voltage) 3~ 3~

1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.