Beijing Laboratory Ultra Pure Water Device
Beijing Laboratory Ultra Pure Water Device Parameters:
name | basic type | Heat source removal type | Low organic matter type | integrative | |||
PRODUCT MODEL | UPL | UPL-UF | UPL-UV | UPL-UVF | |||
application area | Various physical and chemical tests, preparation of solutions, reagents, buffer solutions, student experiments, general chemistry, small-scale biochemical instrument matching, and other routine quantitative and qualitative analyses | PCR analysis, animal and plant cell culture, amino acid analysis, protein purification, electrophoresis/gel analysis, microbial analysis, isotope analysis, etc
| HPLC, IC, ICP, GC/MS, LC/MS, ICP/MS, TOC analysis, trace metal analysis, quantitative analysis of trace elements, preparation of high-purity standard solutions, development of biochips, etc | Molecular biology, life sciences, biotechnology, animal and plant cell culture, genetic research, amino acid analysis, protein purification, toxicology research, drug development, medical testing, and high-precision analysis | |||
Water inlet requirements | Urban tap water: TDS<200ppm, 5-45 ℃, 1.0-4.0kg/cm2 (when the influent TDS>200ppm, it is recommended to choose an external softener) | ||||||
Water output (25 ℃) | 5/10/20/40/60/100 liters/hour | ||||||
Instantaneous water output | 1.5 liters per minute (optional water storage tank) | ||||||
Preprocessing unit | PP filter element, GAC filter element, CTO filter element (external softener is recommended for high hardness source water areas) | ||||||
R. O Reverse Osmosis Unit | Imported original Dow reverse osmosis membrane from the United States | ||||||
Ultra purification system | 4-column fully drooping ultra purification column, imported original resin | ||||||
Dual wavelength UV lamp | / | / | 185/254nm | 185/254nm | |||
Ultrafiltration column | / | Imported original ultrafiltration | / | Imported original ultrafiltration | |||
R. O water index | ≤10μs(25℃) | ||||||
UP Ultra Pure Water Index | 18.2MΩ .cm@25 ℃ | ||||||
Heavy metal ion content | < 0.1 ppb | ||||||
Total Organic Carbon Content (TOC) | / | / | <5 ppb | <5 ppb | |||
Endotoxin (heat source) | / | <0.001 Eu/ml |
| <0.001 Eu/ml | |||
bacterial content |
| <1 cfu/ml | <1 cfu/ml | <1 cfu/ml | |||
Particle content (>0.22 μ m) | / | <1/ml | <1/ml | <1/ml | |||
outlet | 2, R.O. pure water, UP ultra pure water | ||||||
Water quality testing | Online conductivity monitor |
function/characteristic:
Advanced microcomputer control technology enables automatic water production.
Two conductivity meters are used to monitor the water production indicators of reverse osmosis (RO) and ultrapure water (UP) online.
The main components are used to ensure the stability of the instrument and the quality of the produced water.
The front of the chassis is designed with a large front cover, making it easy to replace consumables.
The circuit and waterway design are strictly separated to avoid product circuit failures caused by moisture and water leakage.
● Fully automatic R The O membrane anti scaling cleaning program can be automatically or manually cleaned, extending R O membrane service life.
● Large touch screen LCD, practical and beautiful, easy to operate, displaying parameters such as water quality and temperature of the produced water.
The quantitative water intake function can be operated with just one click, saving time, effort, and peace of mind.
Imported original nuclear grade resin, with a 4-column fully sagging ultra purification column design to ensure * water quality.
Dual wavelength (185/254nm UV lamp component), imported lamp tube, effectively sterilizes and reduces TOC.
Imported original terminal filter (0.45 0.22um) ensures sterile water quality.
●5000DUF ultrafiltration module effectively removes endotoxins (heat source).
The whole machine comes with a one-year warranty and lifetime maintenance.