DuraThe series uses urban tap water (pure water) as the water source, which can conveniently and quickly manufacture secondary waterROReverse osmosis water(DIDeionized water) andUPUltra pure water, using optimized standard two-stage reverse osmosis process: dual pumps, dual membranes, and transition water tank, resulting in lower system energy consumption, less wastewater, and more stable water quality. Even under conditions of poor source water quality, the second levelROThe conductivity of the produced water can remain stable at the same level1-5μs/cm.
existingDura 12/24、Dura 2Sub product series, system water production:12、24rise/Hour (or up to)2.0L/min), Level 2ROReverse osmosis water conductivity1-5μs/cmDeionized water resistivity>16MΩ.cmThe ultra pure water resistivity has reached18.2MΩ.cm, completely in line withGB/T 6682-2008、GB/T33087-2016、ASTM、CAP、CLSI、EPandUSPThe established Grade I water quality standard
Highly integrated
Built in dual water tanks3Pump system with ultra pure water dosing and disinfection function, including3Dual stage reverse osmosis and desktop ultrapure water machine for road water quality monitoring
LCDLarge display screen: size up to106×57mmReal time animated display system operating status and various parameters
Standard dual stage reverse osmosis system, built-in3Pump system, easy to replace overall consumable components
Ultra pure water full pipeline dosing and disinfection function, ultra pure water circulation function,3Road water quality monitoring and alarm function
Comprehensive consumables management function, comprehensive system alarm function
fully automaticROMembrane anti scaling flushing program, timed and quality water intake function
optimizedROMembrane module design, unique4Door opening design
Both pipelines and joints have been obtainedNSFCertification, factory and customer secondary passwords
Powerful pre-processing components, brand new integrated ultra purification components
dual wavelengthUVUV lamp components, imported with original packagingUFUltrafiltration component, double-layer terminal sterilization filter
2、 Technical parameters
Dura12/24series |
Dura 12basic type |
Dura12FHeat removal prototype |
Dura12VLow organic matter type |
Dura12FVintegrative |
|
Dura 24basic type |
Dura24FHeat removal prototype |
Dura24VLow organic matter type |
Dura24FVintegrative |
Pure water production* |
12/24 L/hour |
|||
Ultra pure water production |
up to2.0L/min(When there is water in the water tank) |
|||
Ultra pure water index |
||||
resistivity(25℃) |
18.2MΩ.cm |
|||
total organic carbonTOC* |
<> |
<> |
<> |
<> |
bacteria |
<0.01cfu>0.01cfu> |
|||
particulate(>0.1μm) |
<1>1> |
|||
pyrogen/endotoxin |
N/A |
<0.001eu>0.001eu> |
N/A |
<0.001eu>0.001eu> |
ribonuclease(RNases) |
N/A |
<1pg>1pg> |
N/A |
<1pg>1pg> |
deoxyribonuclease(DNases) |
N/A |
<5pg>5pg> |
N/A |
<5pg>5pg> |
Reverse osmosis pure water index |
||||
Primary reverse osmosis water conductivity |
Conductivity ≈ Raw water conductivity ×5% |
|||
Secondary reverse osmosis water conductivity |
1-5μs/cm |
|||
Source water requirements |
Urban drinking tap water, water temperature5-45℃, water pressure1.0-4.0Kgf/cm2 |
|||
Size and Weight |
Length x Width x Height:545×470×610mm; Weight: Approximately38Kg |
|||
Electrical requirements |
200-240V,50/60Hz |
|||
power |
240W |
|||
system configuration |
host(contain1Purification column set)+Built in water tank2only |
Duraseries |
basic type |
Heat removal prototype |
Low organic matter type |
Dura12FVintegrative |
model |
Dura |
Dura F |
Dura V |
Dura FV |
Ultra pure water production |
up to2.0L/min |
|||
Ultra pure water index |
||||
resistivity(25℃) |
18.2MΩ.cm |
|||
total organic carbonTOC* |
<> |
<> |
<> |
<> |
bacteria |
<0.01cfu>0.01cfu> |
|||
particulate(>0.1μm) |
<1>1> |
|||
pyrogen/endotoxin |
N/A |
<0.001eu>0.001eu> |
N/A |
<0.001eu>0.001eu> |
ribonuclease(RNases) |
N/A |
<1pg>1pg> |
N/A |
<1pg>1pg> |
deoxyribonuclease(DNases) |
N/A |
<5pg>5pg> |
N/A |
<5pg>5pg> |
Deionized pure water index |
||||
resistivity(25℃) |
>16MΩ.cm |
|||
Source water requirements |
Distilled water, deionized water or reverse osmosis water, water temperature5-45℃, water pressure1atm |
|||
Size and Weight |
Length x Width x Height:545×470×610mm; Weight: Approximately25Kg |
|||
Electrical requirements |
200-240V,50/60Hz |
|||
power |
240W |
|||
system configuration |
host(contain1Purification column set) |