Dongguan Saisi Testing Equipment Co., Ltd
Home>Products>Evaluation system for insulation degradation and ion migration characteristics
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
    18926849868
  • Address
    6th Floor, Zone B, Changping Zhigu, Shuishi Road Street, Jiujiang, Changping Town, Dongguan City
Contact Now
Evaluation system for insulation degradation and ion migration characteristics
The evaluation system for insulation degradation and ion migration characteristics is facing the increasingly small, lightweight, and high-density pac
Product details
Details introduction:

Evaluation system for insulation degradation and ion migration characteristicsPurpose:

Faced with the increasingly small, lightweight, and high-density packaging of electronic products, the phenomenon of poor insulation and ion migration caused by factors such as condensation and moisture absorption is becoming more and more prominent. The insulation resistance degradation (ion migration) evaluation system, coupled with a high-temperature and high humidity test box linkage, can achieve high-precision continuous monitoring and efficiently and conveniently evaluate the life and insulation resistance degradation related issues caused by ion migration.

Evaluation system for insulation degradation and ion migration characteristicsApplicable standards:

JPCA-ET04、IPC-TM-650_2.6.3F 、IPC-TM-650_2.6.3.1E、IPC-TM-650_2.6.3.4A 、IPC-TM-650_2.6.3.6。

Evaluation system for insulation degradation and ion migration characteristicsTechnical specifications:


Product Name

Evaluation system for insulation degradation and ion migration characteristics

model

SIR13

120Vsubstrate

250Vsubstrate

500Vsubstrate

1000Vsubstrate

Substrate Testing Department

Test resistance range

320Ω120TΩ

320Ω2.5TΩ

320Ω250TΩ

320Ω500TΩ

320Ω1000TΩ

Number of testing channels

8ch/substrate

16ch/substrate

8ch/substrate

connecting cable

2Root pair/substrate

4Line type

2Root pair/substrate

Load test cable×2

2Root pair/substrate

Load cable×1Test cables×1

Electrical Characteristics

electricity

press

plus

load

ministry

Loading voltage

voltage range1

0.10V120.00V

0.1V250.0V

1.0V500.0V

1.0V1000.0V

voltage range2

0.100V12.000V

--

--

--

Load setting resolution

0.10V/0.001V

0.1V

Basic loading accuracy

±0.3%/FS

±0.3%/FS + 0.5V/FS

Higher output power

96mW/ch

256mW/8ch

300mW/ch

Number of loading groups

1Group(1ch/1Group)

2Group(8ch/1Group)

1Group(8ch/1Group)

Load range

2range

1range

Number of loading channels

1ch

8ch

Large load capacity

2.0μF/1ch

0.47μF/8ch

3300pF/1ch

electricity

press

apparent

show

device

Display range

2range

1range

monitor

scope

voltage range1

0.00V120.00V

0.0V250.0V

0.0500.0V

0.0V1000.0V

voltage range2

0.000V12.000V

--

--

--

Display resolution

0.10V/0.001V

0.1V

Basic display accuracy

±0.3%/FS

±0.3%/FS + 0.5V/FS

Display segmentation unit

1ch

1Group or1ch

1ch

Display cycle

40ms

Number of display channels

8ch

16ch

8ch

electricity

flow

measure

try

Test range

3range

2range

3range

monitor

scope

current range1

0.00μA320.00μA

current range2

0.0000μA3.2000μA

current range3

0.00nA32.000nA

--

0.000nA32.000nA

range setting

320.00μA·3.2000μA

·32.000nA·automatic

320.00μA·3.2000μA·automatic

320.00μA·3.2000μA·32.000nA·automatic

Test smaller resolution

current range1

10nA

10nA

10nA

current range2

100pA

100pA

100pA

current range3

1pA

1pA

1pA

Basic testing accuracy

±0.3%/FS

number of channels

8ch

16ch

8ch

Data collection cycle

regularly30s(Smaller)/During ion migration40ms(Smaller)

Ionic migration testing speed

40ms

Leakage detection

400μs/ch

test cycle

40ms

Other functions

Self diagnosis

External standard resistor diagnosis optional

chain

Test the automatic interruption function when the box door is opened optional

Wire breakage detection

Terminal breakage detection function

--

Temperature and humidity recording in the test tank

Add temperature and humidity testing substrate, with the help of3CS·keylessThe software is large and can be included4Data from a slot optional

Sample temperature recording

pass3CS SMUEach channel can include1a point optional

Large system composition

SIR13

80ch10Substrate)

160ch10Substrate)

80ch10Substrate)

SIR13mini

24ch3Substrate)

48ch3Substrate)

24ch3Substrate)

Control Department

System control computer

Windows XP Pro. SP2suitable forWindows2000Pentium 500MHzabove memory256Mbyteabove

Connection of testing department

GP-IBorEthernet

other

Power outage countermeasures

Save and record data before power outage, and continue recording after power on Uninterrupted power supply must

control unit

structural type

SIR13

SMU 10Slot style

Overall dimensions:W430×H300×D620Excluding protruding parts

Weight: approx.30kgSMU 10When equipped)

Current consumption:5AThe following(100VWhen in use)

SIR13mini

SMU 3Slot style

Overall dimensions:W220×H370×D390Excluding protruding parts

Weight: approx.20kgSMU 3When equipped)

Current consumption:2AThe following(100VWhen in use)

Noise resistance capability

1μspulse2KV 1minute

insulation resistance

DC500V 100MΩabove

Use power supply

AC85V264V 50/60Hz

Usage environment

temperature+10+40 humidity75%RHBelow (without condensation)

Save the environment

temperature-10+ 60

Evaluation system for insulation degradation and ion migration characteristics product image

Online inquiry
  • Contacts
  • Company
  • Telephone
  • Email
  • WeChat
  • Verification Code
  • Message Content

Successful operation!