Can analyze total carbon (TC), total inorganic carbon (TIC), total organic carbon (TOC), soluble organic carbon (DOC), non blowable organic carbon (NPOC), total nitrogen (TN), nitrate nitrogen+nitrite nitrogen (NN), and total Kjeldahl nitrogen (TKN) in water and soil extracts.
The FormacsSERIES series total organic carbon/form nitrogen analyzer provides fast, reliable, and accurate testing of carbon and/or nitrogen for samples from various natural, domestic, and industrial sources. The instrument design meets the latest needs of today's customers, and can detect TOC in clean water such as pharmaceutical and drinking water, as well as analyze wastewater, surface water, seawater, and soil extracts.
The FormacsHT high-temperature combustion catalytic TOC/form nitrogen analyzer adopts a universal modular design and has excellent performance. Used for analyzing total carbon (TC), total inorganic carbon (TIC), total organic carbon (TOC), soluble organic carbon (DOC), blowable organic carbon (POC), non blowable organic carbon (NPOC), and/or total nitrogen (TN). The optional NN reaction module simultaneously measures the values of nitrate nitrogen, nitrite nitrogen (NN), and total Kjeldahl nitrogen (TKN). This unique TKN analysis method is quite fast (5 minutes) and does not require the use of harmful digestive acids.
characteristic:
●Analyze project TC, TIC, TOC, NPOC, DOC, POC, TN, Nitrous nitrogen+nitrite nitrogen and TKN
● High temperature combustion method or low-temperature wet method, NDIR detection (carbon) and chemiluminescence detection (nitrogen)
● 150 position automatic sampler, optional 20 or 40ml EPA-VOA sample bottle
● Automatic pre dilution or post dilution can be applied to samples that exceed the standard, and a series of working standards can be prepared automatically
● Excellent TKN determination method
● Provide automatic stirring, acidification, and gas purging for NPOC analysis
● According to EPA, CEN, DIN, ISO and Standard Methods International Standard Methods
● Compliant with 21 CFR Part 11 specifications
● Scalable measurement of solid TC, TIC and TOC
Model Name: FormacsHT Total Organic Carbon/Total Nitrogen Analyzer
Analysis project: TC, IC, TOC, DOC, NPOC * POC and TN, NN (nitrate nitrogen+nitrite nitrogen), TKN
●Analysis method: High temperature catalytic combustion, with a maximum temperature of 1000 ° C
●Detection method: Synchronous dual wavelength, digital infrared detector (NDIR)
●Applications: drinking water, groundwater, wastewater, seawater, cooling water, surface water, pharmaceutical and industrial products
●Injection: Automatic diaphragm free rotating injection port, built-in sample tubing and injection needle cleaning
●Detectable particle size: Maximum particle size: 450 µ m (optional 800 µ m)
●Measurement range: TOC can reach 60000ppm C, TN can reach 1000ppmN
●Minimum detection limit: TOC 4
ppb C; TN0.01ppmN●
Analysis time: Within 3 minutes for TC/TN or IC analysis●
Injection volume: 0-2500ul electronic micro adjustment syringe
● Dilution: can be pre diluted or post diluted●
Automatic preparation of standard series: can automatically prepare up to 10 points of standard series
● Alarm function: The patented alarm function includes user-defined alarm limits such as carrier airflow, combustion furnace temperature, and Peltier electronic cooler temperature●
Sleep mode: automatic power on and off, unmanned monitoring operation●
Data processing: signal detection: area calculation, peak shape editing function●