1Product Features
1、 Verified operational guidance
2、 High precision
3、 Comprehensive analysis quality assurance functionAQA
2、 Product Introduction
(1) Routine laboratory testing
1、 All colorimetric dishes are equipped with barcodes
2、 automatic recognition10,20and50mmColorimetric dish
3、 Automatically match test range
4、 Automatically storing data for easier data management
5、 built-in200Multiple testing methods, including free ammoniaCO2、SACSpecial testing methods such as chromaticity
6、 Supporting prefabricated reagents to make analysis and testing simpler
7、 User defined method settings
(2) Simplify complex operational steps—There are preset methods or simple self programming
1、 Stepwise data input and operation multi-step testing task: The program provides the required operating parameters, operators, and conditions for each step.
2、 Multi wavelength testing and calculation method
(3) ThephotoLab®7600 UV-VIS OptRFPioneering test free agent testing
1、 Just put the sample into a quartz colorimetric dish, select the method, and you can directly read it!7600 UV-VISCan be achieved throughOptRFDirect detection of complex water samples in the ultraviolet spectral region using the (reagent free) measurement methodCODThe concentration of nitrate nitrogen and nitrite nitrogen is easy and fast to operate. Suitable for rapid verification and sample retention of reported data.
2、 Real time reading
3、 User calibration program makes data more accurate
4、 Economy, environmental protection, and harmlessness
5、 supportIQ Sensor NetGood combination with photometry
(4) Dynamics—Chemical reaction rate
1、 Measurement at room temperature
2、 Can store user data
(5) Quality control and R&D spectra can save user spectra and data processing
1、 Classify spectra using different colors
2、 Spectral editing (summation, comparison, etc.)
3、 Readable absorbance and transmittance
3、 Technical parameters
Instrument Name |
photoLab®7100 VIS |
photoLab®7600 UV-VIS |
wavelength range |
320 ~ 1100 nm |
190 ~1100 nm |
raster |
Equipped with stepper motor and reference beam grating monochromator |
|
monitor |
7”Background light color display |
|
bandwidth |
4 nm |
|
light source |
tungsten lamp |
xenon lamp |
Test mode |
Concentration, absorbance,%Transmittance, dynamic analysis, spectral curve scanning or%Transmittance, multi wavelength stepwise testing. |
|
scanning speed |
700 ~ 2000 nm/min |
|
wavelength accuracy/repeatability |
± 1 nm / < 0.5 nm |
|
Colorimetric accuracy/repeatability |
- 0.003 E(E < 0.600)0.5%(0.600 < E < 2.000) |
|
Colorimetric dish |
Automatically identify different types of colorimetric dishes, with optional outer diameters16mmrotundity10mmSquare20mmRectangle50mmRectangular colorimetric dish |
|
bar code |
Automatically identify the testing range of all colorimetric dishes |
|
data storage |
5000A measurement value,500Spectral analysis,400A dynamic curve. |
|
Method and Curve |
200Multiple built-in testing methods,1000A custom method,20Various programs are available for selecting dynamic and spectral curves. |
|
interface/update |
1 individualUSB-A. 1individualUSB-B. 1individualEthernet/Through the internet andUSBInterface update |
|
enclosure ratings |
IP 30 |
|
power supply |
Universal power supply, can also be powered by car batteries using standard adapters and cables |
|
temperature range |
operation temperature +10 °C ~35 °CStorage temperature -25 °C ~65 °C |
|
weight/size |
make an appointment4.5 kg/404 mm x 197 mm x 314 mm |
|
attachment |
PCThe software includesphotolab®database+ PhotoLab®Color library, on-site situation,AQAThe inspection tool |