Beijing Huayi Tongtai Technology Co., Ltd
Home>Products>German WTWTurb2000 online turbidity measuring instrument
German WTWTurb2000 online turbidity measuring instrument
The smaller sampling pool of the German WTW online turbidity measuring instrument Turb 2000 reduces calibration costs and provides response speed; The
Product details

Beijing Huayi Tongtai Technology Co., Ltd. specializes in the sales of imported analytical and testing instruments. We provide a wide range of products, competitive prices, and excellent after-sales service.

GermanyWTWOnline turbidity measuring instrumentTurb 2000A smaller sampling pool reduces calibration costs and provides response speed; GermanyWTWOnline turbidity measuring instrumentTurb 2000Simple operation and maintenanceIt can be almost beautifully used in the laboratory.
GermanyWTWOnline turbidity measuring instrumentTurb 2000Product Introduction:
measuring principle
90°WLScattered light method
Measurement is conducted in accordance withUSEPA180.1Standardized90°Scattering light method.
The turbidity of the medium is determined by the intensity of scattered light. The white light beam is scattered by solid particles in the medium, and the scattered light is mixed with the white light source90°Arrange scattered light detection units for detection.
90°NIRScattered light method
Measurement is conducted in accordance withISO7027/EN27027Standardized90°Scattering light method. The turbidity of the medium is determined by the intensity of scattered light. Near infrared light of a certain wavelength is scattered by solid particles in the medium, and scattered light is mixed with an infrared light source90°Arrange scattered light detection units for detection.

function
BacklightLCD
BacklightLCDEnsure clear display in dark conditions. Backlight is suitable for continuous operation. The brightness can be adjusted in the menu.
Water vapor elimination
MicroTOLEquipped with a continuous water vapor elimination system. The replaceable drying bag is installed under the instrument panel for drying the air. The system heat is used to heat the air, and the fan inside the instrument circulates the heated air around the sample cell, eliminating the process of drying the sample cell.
call the police
provide2An independent and programmable alarm relay.
ultrasonic cleaning
This option is used for continuous cleaning of the sample tank. It cannot completely replace manual cleaning, but this system will greatly extend the cleaning cycle.
Attention: The system requires the use of a special sample pool.
correct
The optional calibration kit is a completely sealed test tube, similar to laboratory calibration steps. This dry calibration method is very fast, clean, and reusable. The menu on the screen will guide you to quickly and easily calibrate.

System connection installation

Wall mounted installation Suggested for indoor use,Pay attention to the following items during installation:
nEnsure that the temperature does not exceed a large allowable operating temperature range(0 ~ 50℃
nLeave space at the top of the instrument panel200mmEnsure sufficient space for calibration and maintenance of the sample pool
nThe instrument should be installed as close as possible to the sampling point(2~3m)To ensure a quick response.

GermanyWTWOnline turbidity measuring instrumentTurb 2000productcharacteristic
Can provide white light source type or infrared light source type
Quick and easy calibration
A smaller sampling pool reduces calibration costs and provides response speed
Automatic continuous ultrasonic cleaning extends the cleaning cycle
Simple modular design
Simple operation and maintenance
Low price
• RS-485 Modbusinterface
integration
Optional calibration toolkit

GermanyWTWOnline turbidity measuring instrumentTurb 2000Technical Parameter

measuring range

0 – 1000.0 NTU
0 – 100 NTU

measuring principle

90Scattered light

accuracy

±2%Reading or±0.02 NTUTake the larger of the two (less than)40 NTU
±5%Reading (greater than)40 NTU

resolution

0.0001 NTU (less than10 NTU)selectable

reaction time

5-500Seconds, adjustable

display

Multi line LCD backlit display

light source

white light– 7Annual lifespan
infrared light– 10Annual lifespan

call the police

2A programmable relay,120-240VAC/2A

Analog Output

Powered4-20mA600ΩA large load

Port

two-wayRS-485Modbus

High water pressure

Built in pressure regulating valve, with high pressure1380KPa

velocity of flow

100ml/divide~ 1rise/divide

Sample temperature range

+1 ~ +50℃

Liquid bonding material

Nylon, silicate glass, silicone, polypropylene, stainless steel

power supply

100-240VAC47-63Hz80VA

Insulation level

Double layer insulation, anti pollution level2Level, overvoltage protectionIIclass

environmental condition

Not recommended for outdoor use, high altitude2000Rice, high humidity95%RH(Non condensing)

degrees of protection provided by enclosure

IP66/NEMA 4X

Certification and Rules

White light type meetsU.S.EPA 180.1
Infrared type meetsISO 7027
CEauthentication,ETL UL61010B-1ETLc CSA 22.2 No. 101-1-92

External dimensions

35 cm x 30 cm x 30 cm

Transport weight

2.5Kg