- Product Description
-
Grant OptinmaTMThe TC120 series is a basic constant temperature controller that is easy to operate. With just one knob and two function keys, temperature settings and related operations can be quickly set. TC120 can control the temperature range from -20 ℃ to 120 ℃. The TC120 series thermostatic control can be used in conjunction with GRANT ST and P series baths. When applied to low-temperature control, a refrigeration device is required.
Functional features:Optima ™ Digital constant temperature controller, intelligent control optimization ™ ("ICO") , Adaptive intelligent PID temperature control
The temperature range for heating and cooling is -20 ℃ to+100 ℃
The temperature stability is ± 0.05 ° C
Bath capacity can be selected with different materials (stainless steel material 5L, 12L, 18L, 26L, 38L; plastic material 5L, 12L, 18L)
Simple and convenient programming functions
Four preset temperatures can be set for easy retrieval
TC120 is equipped with a built-in circulation pump, over temperature protection device, timing function, and sound alarm function
Product Features:Notify users in a timely manner when there are abnormal situations with sound and light alarms, and TC120 also has sound alarm function
Rotary control key with 2 buttons - allows for quick and accurate settings
The timing function facilitates users to conduct timed experiments (TC120)
User calibration function - facilitates maximum accuracy at the desired operating temperature
High power built-in circulation pump - can provide constant temperature liquid (TC120) for external devices
The bridge type platform facilitates the movement of instruments - making the use of the constant temperature controller convenient and intuitive, and making full use of space.
Clear digital display - temperature can be accurately seen even from a distance
1 operation setting temperature and 3 adjustable preset temperatures make operation more convenient
Optional detachable bath cover - reduces liquid evaporation and prevents sample contamination, insulated handle
Sturdy structure - made of stainless steel material, with good corrosion resistance and durability
Grant Optima ™Technical parameters of constant temperature controller ●=Standard configuration Universal digital type High performance digital type T100 TC120 TX150 TXF200 Stability (DIN12876) @ 70 ℃ ℃ ±0.05 ±0.05 ±0.01 ±0.01 Uniformity (DIN12876) @ 70 ℃ ℃ ±0.1 ±0.1 ±0.05 ±0.05 Set accuracy ℃ 0.1 0.1 0.1 (0.01 using Labwise) ™ Software) Display mode 4-digit LED Color screen QVGA TFT Timer function - 1 to 9999 mins 1min to 99h59mins Number of preset temperatures 3 3 3 3 Two-point calibration ● ● ● ● Temperature compensation function - - ● ● External temperature probes (TXPEP, TXSEP) - - ● ● communication interface - - USB&RS232 USB&RS232 Programming mode - - PC remote control/1 programming 30 steps Interface control/remote PC control/10 programming 100 steps relay - - 1 1 Security Over temperature protection fixed Adjustable disconnect Low liquid level protection - floating on light ● ● ● ● Alarm (can set a relay control switch) - - High and low temperature High and low temperature heating power 230V KW 1.3 1.3 1.9 1.9 120V KW 1.4 1.4 1.4 1.4 Total power 230V KW 1.4(50-60Hz) 1.4(50Hz) 2.0(50Hz) 2.0(50-60Hz) 120V KW 1.5(50-60Hz) 1.5(50-60Hz) 1.5(50-60Hz) 1.5(50-60Hz) Above the height of the edge of the bath mm 200 200 200 200 Depth below the edge of the bath mm 135 135 135 135 Grant Optima ™ Thermostatic circulation pump (built-in) Maximum pressure Water bar - 210 310 530 Maximum flow velocity Water L/min - 16 18 23 (adjustable flow rate) Pipeline diameter In/out mm - 6/11 6/11 6/11 Grant immersion thermostats can be paired with Grant stainless steel and plastic baths. If equipped with K-type clamps, they can also be used in vertical baths with a maximum wall thickness of 35mm or circular baths with a maximum wall thickness of 30mm (diameter 300mm). The maximum volume can reach 50 liters, and the highest and lowest temperatures that can be achieved are determined by the insulation capacity of the bath and the cooling and heating capacity of the cooling and heating equipment.
ST series stainless steel and P-series plastic bath Volume (L) External bath size Working size (L × W) Min/Max liquid level depth Internal bath size (L × W × H) External dimensions (including controller) (L × W × H) ST5-5L stainless steel 150×150mm 85/140 mm 350×140×150 mm 330×180×395 mm ST12-12L stainless steel 205×300mm 85/140 mm 325×300×150 mm 360×330×395 mm ST18-18L stainless steel 385×300mm 75/130**mm 505×300×150 mm 540×330×395 mm ST26-26L stainless steel 385×300mm 125/180**mm
505×300×200 mm 540×330×405 mm ST38-38L stainless steel 575×300mm 125/180**mm 690×300×200 mm 730×333×450 mm P5-5L plastic 120×150mm 85/140mm 240×160×150 mm 390×200×380 mm P5-12L plastic 210×280mm 85/140mm 325×280×150 mm 415×350×380 mm P18-18L plastic 280×325mm 85/140mm 510×290×150 mm 600×350×360 mm parts Bath cover*
Reduce heat loss and bathing
Liquid evaporation and avoidance
contaminatedPolypropylene ball*
(Recommended packaging quantity)
Sample rack?
Fill the bath capacity
Split utilization (recommended configuration sample)
Quantity of shelvesRaise the sample rack
Raise the sample height to suit
Liquid levelOptional configuration of cooling system**
By using refrigeration coils to keep the system temperature lower than the ambient temperature, it will also have an impact on the temperature of the working environment
Reduce to the lowest levelImmersion cooler
Control the temperature through a bath temperature controller, and connect the refrigeration coil of the refrigeration unit through a hose to absorb heat and cool
Heat exchange coil
Design supporting refrigerants or systems
Use of cold cycle unitC1G (0 to 40°C***) C2G (- 15 to 40°C***) CW5 (2 ° C higher than the refrigerant) STL5Flat bottomed, stainless steel
1×PS20 1×QR - YES - YES STL12Yamagata, hinge (removable) stainless steel
1×PS20 2×VR RS14 YES - YES STL26Yamagata, hinge (removable) stainless steel
2×PS20 4×VR RS22 YES - YES STL26Yamagata, hinge (removable) stainless steel
2×PS20 4×VR RS28 YES YES YES STL38Yamagata, hinge (removable) stainless steel
3×PS20 6×VR RS28 or RS38 YES YES YES PL5Horizontal, made of stainless steel material
1×PS20 1×QR - - - - PL12Bending, made of plastic material
1×PS20 2×VR RS14 - - - PL18Bending, made of plastic material
2×PS20 4×VR RS22 - - - *When the operating temperature is between 60 ° C and 100 ° C or below room temperature, a bath cover or polypropylene ball should be used. If it is above 100 ° C, a bath cover must be used.
**When the cooler is turned off, the cooling coil can still be placed in liquid above 100 ℃, and can cool the liquid at 100 ℃, but it will stop working when it is below 40 ℃
***The minimum working temperature (without cooling device) is 5 ℃ above room temperature (S5 bath is 15 ℃ above room temperature)
Sample rack specifications (number of test tubes per rack)VR stand Test tube model capacity QR bracket Test tube model capacity VR-13 ?10-13 mm 65 QR-13 ?10-13 mm 30 VR-19 ?16-19 mm 36 QR-19 ?10-13 mm 16 VR-24 ?24 mm 23 QR-24 ?10-13 mm 10 VR-30 ?30 mm 14 QR-30 ?10-13 mm 5 VR-SE 0.5ml 102 QR-SE 0.5ml 44 VR-LE 1.5ml 75 QR-LE 1.5ml 35 Grant supporting refrigeration system Immersion cooler Heat exchange coil C1G C2G CW5 Cooling power @20℃ W 350 400 - @0℃ W 110 320 - @-10℃ W - 170 - Total power VA 300 500 - size D/W/H mm 460/305/225 - Hose length L mm 925 925 - pipe diameter ?/I mm 77/55 77/55 77/55 Pipeline diameter inlet/outlet mm - - 7 Power Supply 120V(60Hz) or 230V(50Hz) - *The optional VTP pump will bring additional heat to the bath, so if refrigeration accessories are not provided, the minimum temperature that the bath can reach will increase. Attention: When ordering a VTP pump, please indicate the model of the Grant bath provided.