Product Introduction
The HXDCL-400 residual chlorine/chlorine dioxide sensor adopts advanced constant voltage principle and is used to measure residual chlorine/chlorine dioxide in water bodies. This method utilizes a stable potential applied between the polarizing electrode and the reference electrode, and different measured components generate different current intensities at this potential. The instrument calculates the concentration of the measured component by collecting and analyzing the current signal.
Product Features
The HXDCL-400 residual chlorine/chlorine dioxide electrode has a simple structure, is easy to clean and replace, and does not require the replacement of membranes and reagents during electrode use. It is easy to maintain and ensures the stability, reliability, and accuracy of the instrument's long-term operation.
Typical applications
Used for measuring residual chlorine and chlorine dioxide in wastewater and water treatment from water plants, drinking water distribution networks, swimming pools, hospitals, etc.
Technical Parameter
Original texture
|
Constant voltage method
|
measuring range
|
0-20 mg/L
|
Schoolaccurate
|
Two point correction
|
Sensitivity
|
78%--155%
|
temperature compensation
|
Manual or automatic temperature compensation
|
temperature coefficient
|
2%/℃
|
Electricitysource
|
AC power supply: 100~240VAC 5W
DC power supply: 9~36VDC5W
|
Losinggo out
|
Two way isolation 4-20mA
|
Communication protocol
|
MODBUS RS485
|
Flow rate requirements
|
30-90L/h (60-90L/h is the best)
|
operation temperature
|
0 to 50 ℃
|
electrode size
|
Diameter 13mm * Length 120mm
|
Electrode material
|
316L stainless steel shell
|
Cable Length
|
Standard 3-meter cable
|