The HQ11D digital pH analyzer can be used with any standard pH IntelliCAL electrode from Hash Company. HQ11D uses a lot of advanced software and technology, which is a very economical measurement method. Calibration data and calibration history can be carried along with each IntelliCAL electrode, allowing operators to switch calibrated electrodes between analyzers. The electrodes should be calibrated in a controllable environment and can be used on-site. Automatic detection of standard procedures and calibration reminders can ensure the accuracy of data. It can store a total of 500 data records by linking each measured sample ID, user ID, and electrode serial number.
Technical Parameter
Please refer to the technical parameters of the pH electrode for the measurement range
Power supply: 4AA battery; Battery life:>200 hours with AC/DC power connection included
Language: Optional: English, French, German, Italian, and Spanish
Single parameter display: can display 1 reading.
Data memory: 500 sets of data
Data storage: Calibration data is stored in logs. It can be automatically stored in the "press to read" mode and interval measurement mode. Manual storage is required in the 'continuous reading' mode.
Temperature correction/compensation: automatic shutdown
Lock display function: Continuous measurement mode or press to read measurement mode (with numerical averaging function).
Automatic recognition standard: There are three options to choose from:
Color classification: 4 7 10pH
IUPAC: 1.679 4.005 7.000 10.012
DIN: 1.09 4.65 9.23
Waterproof: The measuring instrument casing can be soaked in water up to 1 meter deep for 30 minutes (IP67)
Technical indicators:
Electrode connection port:pH, conductivity,O2
Ph:
Range: 0-14
Resolution (optional):0.1/0.01/0.001
Accuracy:±0.002
Temperature compensation: automatic
ORP:
Range:±1500mV
resolving power:0.1mV
Accuracy:±0.1mV
temperature
Range:–10~+110℃
resolving power:0.1℃
Accuracy:±0.3℃
Automatic reading: Yes
Automatic calibration: PH:S; HQ30/40D
Correction Point/Calibration and Verification: Standards/Reminders: at most4Points
Analyzer interface:Equipped with waterproof functionUSBInterface, can be used for printers, computers, keyboards, andUSBflash memory
(ExceptHQ40dOther models require optional selection
Password Protection: Yes
Data management:GLPData management methods
Data storage: 500Individual readings; Can manually or automatically store data
sampleIDAnd the operatorID:Arabic numerals: maximum12One character;12Name of the sample and20User Name
Automatically record sample serial number(0-999)
Measurement mode:Manual, intermittent, continuous; Analysis method editable
Display:Backlit graphic display;240×160Pixel; Automatic shutdown power-saving mode; Can display time and date
Power Supply:115V-250Vperhaps4jointAAbattery
Protection level:IP67
Size and weight:95×197×36mm(H×W×L),323g(No battery)