Immersion installation of bracket (pool wall)
|
²fixed base Insert four M10 steel expansion bolts at appropriate positions on the pool wall. ²Connection (as shown in the left figure) Thread the cable of the electrode through the connecting sleeve, installation pipe, sleeve, pipe cover, and waterproof head in sequence; Connect the thread of the electrode to the threaded hole of the connecting sleeve, tighten it, and seal it;The stainless steel connecting sleeve is connected to the installation pipe clamp; ² fasten Thread the installation pipe through the sleeve, and fix the sleeve in the appropriate position on the installation pipe with two locking screws based on the measured depth on site; Tighten the pipe cover and installation pipe; Tighten the waterproof head, secure the cable, and prevent rainwater and other substances from entering the installation pipe. ² install Insert the sleeve of the installed electrode installation tube into the bracket bracket installed on the pool wall and fix it with two M6 bolts. AttentionWhen connecting the electrode to the connecting tube, please rotate the connecting tube instead of the electrode, otherwise the cable of the electrode may be damaged Please rotate the installation pipe when connecting it to the connecting pipe. |
1. Base 2, bracket support3waterproof head 4. Pipe cover5. Connecting sleeve locking screw M6 × 10 6casing 7Installation pipe8. Card equipment9. Electrode (with connecting sleeve)