Karl Fischer Moisture Analyzer - High Temperature Special Karl Fischer Furnace ADP-512S
Karl Fischer Moisture Analyzer - High Temperature Special Karl Fischer Furnace ADP-512S Standard Method:
ISO 2596 Iron ores - Determination of hygroscopic moisture in analytical samples - Gravimetric, Karl Fischer and mass-loss methods.
ISO 7335 Iron ores - Determination of combined water content - Karl Fischer titrimetric method.
JIS M8211 Iron Ore - Quantitative Method for Chemical Water
Heat the sample at 105 ° C in a nitrogen gas stream to remove hygroscopic water, then continue heating at 950 ° C and use ethylene glycol to absorb the free bound water. Perform electrical titration using Karl Fischer reagent solution.
Karl Fischer Moisture Analyzer - High Temperature Special Karl Fischer Furnace ADP-512S Main Features:
Suitable for moisture determination of solid and powder samples such as metal powders and ceramics.
Karl Fischer Moisture Analyzer - High Temperature Special Karl Fischer Furnace ADP-512S Technical Parameters:
Heating temperature range: 50-1000 ° C.
Temperature control method: PID temperature control.
Traffic display: 100-300mL/mim.
Heating tube: Quartz glass.
Sample tank: Quartz glass.
Sample capacity: 10mL.
Size: 835 (W) × 340 (D) × 345 (H) mm.
Weight: Approximately 30kg.