Large scale alternating humidity and heat test equipment:
The walk-in constant temperature and humidity laboratory is suitable for aerospace products, information electronic instruments and meters, materials, electrical and electronic products, and various electronic components to experience their various performance indicators in high and low temperature or humid and hot environments.
、
implement |
Nickel chromium alloy electric wire heater |
|
|||||||
Humidifier |
Surface evaporation type |
|
|||||||
Blower fan |
Centrifugal fan |
|
|||||||
Airflow mode |
Broadband forced airflow circulation (top out and bottom in) |
|
|||||||
Condensation method |
Air cooling |
|
|||||||
performance |
temperature range |
- 20℃~+150℃;- 40℃ ~ +150℃;- 70 ℃~+150 ℃ (customizable according to requirements) |
|||||||
Humidity range |
20%~98%RH |
|
|||||||
Temperature fluctuation |
±0.5℃ |
|
|||||||
Humidity fluctuation |
±2%RH |
|
|||||||
Temperature uniformity |
±1.5℃ |
|
|||||||
Humidity uniformity |
±3%RH |
|
|||||||
Heating time |
3 ℃/min * Fast up to 10 ℃/min |
|
|||||||
Cooling time |
1 ℃/min * Fast up to 5 ℃/min |
|
|||||||
Box material |
Inner box: # 304 mirror stainless steel plate |
Outer box: # 304 stainless steel or steel plate spray coated |
|||||||
Insulation material |
Hard polyurethane foam+glass wool |
|
|||||||
controller |
Original imported Korean TEMI880 human-machine interface touch screen controller or TEMI300 LCD button controller |
|
|||||||
Freezing system |
French Taikang fully enclosed compressor/environmentally friendly refrigerant |
|
|||||||
Protector |
Compressor overheating, overcurrent, overpressure, heating humidification air burning, box temperature rise, water shortage alarm system |
|
|||||||
Power Supply |
3 ф 5W380VAC ± 10% or 1 ф 3W220VAC ± 10% |
|
|||||||
Standard configuration |
One observation window (three-layer insulated tempered glass); One ф 50mm test hole (located on the left side); Two sets of sample racks (adjustable height); One fluorescent lamp inside the box; Four mobile work wheels; One humidification water tank; Communication interface; Connect a set of computer operating software. |
|
|||||||
Optional parts |
Recorder |

Technical parameters:
The performance measured after stabilizing for 2 hours without test load or layer frame. In addition, the temperature rise and fall time refers to the performance of the air-cooled type at room temperature+25 ℃ and no load
1) Temperature range:- 20-80℃(T);- 40-80℃(F);- 70-80 ℃ (S) high temperature can be achieved up to 120 ℃ according to the user's needs.
2) Humidity range: 35-95% RH
3) Uniformity: ① Temperature: ≤ 2.0 ℃ ② Humidity: ≤ 5%RH
4) Control accuracy: ① Temperature ± 2 ℃ ② Humidity ± 2.5% RH
5) Heating and cooling time: ① Heating up at about 3.0-4.0 ℃ per minute (expandable to 2-10 ℃/min) ② Cooling down at about 0.7-1.2 ℃ per minute (expandable to 2-10 ℃/min)
6) Accuracy range: ① Temperature: Setting accuracy ± 0.1 ℃, indication accuracy ± 0.1 ℃, resolution ± 0.1 ℃; ② Humidity: Setting accuracy ± 1% R H, Indication accuracy ± 1% R H, Resolution ± 1% R H
7) Internal box size: Made according to user needs, with volumes ranging from 1 cubic meter to 50 cubic meters, all of which can be made, and even larger ones.

Imported programmable touch LCD Chinese dialogue display, microcomputer integrated controller
The temperature and humidity display control instrument adopts a fully imported large screen (5.7-inch LED display) from Japan, with simple screen operation, easy program editing, no need for button input, and direct touch options on the screen. Accuracy: 0.1 ℃ (display range)
Resolution: ± 0.1 ℃;
Temperature sensor: PT100 platinum resistance thermometer;
Control mode: thermal balance temperature and humidity control mode; All electrical appliances use Schneider series products
Temperature and humidity control adopts P I. D+S.R system coordinated control on the same channel
Equipped with automatic calculation function, it can immediately correct the temperature and humidity change conditions, making temperature and humidity control more accurate and stable
The controller operation interface is available in both Chinese and English, and the real-time operation curve chart can be displayed on the screen
It has a capacity of 10 programs, 100 segments per group, and 999 steps per segment that can be cycled. The maximum time setting for each segment is 99 hours and 59 minutes

characteristic:
1. We can provide temperature and humidity environment testing for large parts, semi-finished products, and finished products in the industry
Mainly composed of control panel, distribution panel, distribution panel, insulation storage board, air supply board, heater, humidifier, and cooling system
Composed of a freezer.
2. Combination of color storage board units, the inner cellar can be enlarged freely and disassembled easily, and can be designed and matched according to customer size requirements
Relocation of factory expansion.
3. Adopting a horizontal diffusion vertical heat exchange circulation system with multi wing centrifugal air supply method to avoid any dead corners that may cause damage to the indoor environment
The temperature and humidity are evenly distributed.
4. The freezer adopts imported compressors from Europe and America, and uses environmentally friendly refrigerants. The refrigeration system adopts a binary low-temperature system
Design the circuit system and use different compressors in different temperature ranges to extend the service life of the equipment.
5. The control system can be connected to the computer through an RS-232 interface, and users can design programs on the computer screen
Collecting test data and records, calling program execution, remote control of machine switches, and other functions.
When an abnormal situation occurs, the controller screen will automatically display the fault status in real time, cut off the power switch, and lift the alarm
Provide troubleshooting methods.
7. It has the function of timing marker contacts, which can output control contacts to the outside during operation, providing users with access
Conduct dynamic testing and computer monitoring.
8. The material of the storage board is made of stainless steel and heat-resistant steel plate baked paint, with a sturdy structure, waterproof and beautiful appearance.
9. Temperature and humidity laboratory,Temperature and humidity experimentroomControl method and features: Balanced temperature control systemBTHC Special
Specially designed ceiling fan forced air supplyCircular system, with P I. D. Control SSR through methods,Heating the system
The humidification amount is equal to the heat and moisture loss, so it can last for a long timeStable use during the period.
What industry standards does the alternating humidity testing machine meet?
Answer: Meet the standards |
GB10586-2006 Technical Conditions for Damp Heat Test Chamber |
GB10589-2006 Technical Conditions for Low Temperature Test Chamber |
|
GB10592-2006 Technical Conditions for High and Low Temperature Test Chambers |
|
GB11158-2006 Technical Conditions for High Temperature Test Chamber |
|
GB/T2423.1-2001 (IEC68-2-1) Test A: Low temperature test method |
|
GB2423.2-2001 (IEC68-2-2) Test B: High temperature test method. |
|
GB2423.3-93 (IEC68-2-3) Test Method for Ca Constant Damp Heat Test. |
|
GJB360.8-87 (MIL-STD-202F) High temperature life test. |
|
GJBl50.3-86 (MIL-STD-810D) High temperature test method. |
|
GJBl50.4-86 (MIL-STD-810D) Low temperature test method. |
|
GJBl50.9-93 (MIL-STD-810D) Wet Heat Test Method. |
|
GB/T 2423.3-1993 Test Ca: Constant Damp Heat Test (Damp Heat Type Only) |
|
GJB 150.9-1986 Damp Heat Test (Damp Heat Type Only) |
|
GB/T5170.2-1996 Basic parameter calibration methods for environmental testing equipment for electrical and electronic products Temperature testing equipment |
|
GB/T5170.5-1996 Basic parameter calibration methods for environmental testing equipment for electrical and electronic products - Damp heat testing equipment |
4Installation site
In order to facilitate heat dissipation and maintenance of the enclosure, the location where this equipment is installed must meet the following conditions:
1. The distance between adjacent walls or objects should have the space shown in Figure 1.
Company:mm
1. In order to stably utilize the functions and performance of the test chamber, a place with a perennial temperature of 15-25 ℃ and a relative humidity of no more than 85% should be selected.
2. The environmental temperature of the installation site should avoid sudden changes.
3. It should be installed on a horizontal surface.
4. It should be installed in a place without direct sunlight.
5. It should be installed in a well ventilated area.
6. It should be installed in a place far away from combustibles, explosives, and high-temperature heat sources.
7. It should be installed in a place with less dust.
8. Install as close to the power supply as possible.
1Electrical installation
1. Pay attention to the power capacity. Do not use the same power supply for multiple machines at the same time to avoid voltage drop, affecting machine performance, and even causing malfunction and shutdown. Please use a dedicated circuit.
2. The variation of power supply voltage should not exceed ± 5% of the voltage (maximum allowable voltage is ± 10%)
3. If it is a three-phase power supply, please pay attention to phase loss protection. If it is confirmed that the three-phase power supply is energized but the machine is not operating, then the machine may be in reverse phase, and only the adjacent two power lines need to be exchanged with each other
There is a grounding terminal at the rear of this machine, please ground it. The grounding wire must be constructed properly. According to the technical guidelines for electrical equipment, for the third type of grounding project, the grounding resistance should be maintained atBelow 50 Ω.
4. If you connect the grounding wire to the water pipe, the water pipe must be a grounded metal pipe. Not all metal pipes can be effectively grounded
5. Do not connect the grounding wire to oil or gas pipes.
6. Do not connect the power supply used for the test product to the power supply of this machine, as it has already been planned and designed. Adding other loads may result in excessive load.
7. Is there a possibility of damaging wiring and piping during installation.
6、 Water tank inlet
The water adding device of this equipment is automatically controlled
7、 Structure Introduction
This test box breaks through the design concept of existing domestic products, with a beautiful appearance, ideal assembly method, compact box structure, and extremely convenient installation and use.
The outer shell of this test box is made of cold-rolled steel plate surface spray treated, with smooth lines and beautiful appearance. The inner chamber workshop adopts stainless steel brushed plate, with a smooth and shiny surface. The rear of the workshop is equipped with a blower duct, which contains heaters, refrigeration evaporators and other components. It is also equipped with a blower fan to blow cold and hot air into the workshop to ensure the required temperature.
The refrigeration system of this test chamber is installed at the lower part of the box, and the refrigeration machine adopts a French Taikang fully enclosed compressor, which is easy to install, reliable in performance, and low in noise.
The insulation material adopts ultra-fine glass wool, which has good insulation effect.
This test chamber is equipped with two silicone rubber door seals, which have good sealing performance.
8、 Performance characteristics
1. Studio size: 500 * 600 * 750 (depth x width x height) mm
2. Temperature range:- 60℃~+150℃
3. Temperature uniformity: ± 2 ℃
4. Temperature fluctuation: ± 0.5 ℃
5. Humidity range: 20-98% R · H
6. Humidity deviation:+2~-3% R · H
7. Cooling rate: 0.7-1 ℃/min (unloaded)
8. Heating rate: 2-3 ℃/min (no load)
9. Noise (dB): ≤ sixty-five
10. Power supply: AC3800V 50HZ
11. Power: 5.5KW
1. Inspection before operation
1. Confirmation of power supply and grounding wire
Is the power supply properly connected according to specifications and properly grounded.
2. Check the over temperature protector
Check if the over temperature protector is set
3. Confirmation of water
Is there enough water stored in the water tank?
4. Confirmation of water pipes
Is the drainage pipe properly connected?
5. Confirmation of wet bulb test gauze
lIs it clean (clean every seven days, everyReplace the test gauze once a month)
Test whether the gauze is moist (when there is water in the water supply tank and the water level is normal, but the gauze cannot be wetted, please replace it with a new test gauze).
6. Confirmation of humidification level of humidifier
lIs the water level normal. After the humidifier is supplied with water3-5 minutes, check water level)
Is the water used in the humidifier clean? (The humidification level of the humidifier should be regularly drained and cleaned with a brush to maintain cleanliness.)
7. Check if the over temperature hold switch is set correctly.
Note: When only conducting temperature tests, the test cloth should be removed. If the test cloth isAfter operating at high temperatures above 85 ℃, the test cloth should be replaced before the next operation, otherwise it may not be able to absorb water again. (When testing the packaging of the cloth, it has undergone * * processing.)
1. Usage method
1. Connect the power supply and turn on the main power switch
2. Inject the experimental water into the water storage tank and observe the water level of the humidifier to see if the water has submerged the heating tube
3. Hang the prepared wet ball gauze on the wet ball sensor, with a length of both sides placed in the wet ball sink. (Soak the wet ball gauze before hanging)
4. The water level in the wet bulb tank is adjusted by a level switch, which is called water balance adjustment. If the water level in the wet bulb tank is too high in the working room, the level switch will automatically adjust, otherwise it will be raised
5. Press the power switch, the indicator light will turn on, and set the temperature, humidity, and time to the desired level (operate according to the instrument manual)
6. Inspection and testing operation
7. Make necessary test records according to the test requirements
8. After the natural shutdown of the test according to the prescribed time, the power should be turned off in a timely manner to make all systems in the test box in a static state.