Beijing Feistek Technology Co., Ltd
Home>Products>Laser Raman Spectroscopy Online System PI-200-I
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
  • Address
    Room 1202, Unit 2, Building 8, Poly Oriental Center, No. 129 Beili, Balizhuang, Chaoyang District, Beijing
Contact Now
Laser Raman Spectroscopy Online System PI-200-I
Process Instruction Series
Product details

Process Instruments is a professional company specialized in researching applications of laser Raman spectroscopy analysis. Complete research on customer samples includes collecting spectra from samples, establishing a Chemometric model, building and maintaining a customer model library, determining the most suitable laser spectral wavelength, sample structure Prepare for specific applications based on requirements such as CCD detectors.

We manufacture an effective, concise, and sensitive (f/2.0) spectrometer and a frequency stable, narrow spacing, and long-life laser (design life>6 years).

Our Raman probes with fixed laser filters>10-8 and top mounted interface filters can be used with 1000 meter long optical fibers. They can withstand 400 ℃ and 200 kg pressure when directly inserted into pipelines. We provide an efficient multi-channel laser Raman analyzer that can accommodate 18 or 9 channels for individual control. The software automatically loads appropriate modes for various channels and sample streams.

Our PROSpectTM Raman software is compatible with GRAMS/AI7TM and can be used in both laboratory and online tuning process control. The software features include automatic neon lamp absolute wavelength calibration, real-time chemometrics, and continuous laser wavelength tracking to ensure stable stoichiometric operation. All spectral data is calibrated to ensure that the model can be easily transferred from one Raman instrument to another. Our online multi-channel system (model PI-200-I) includes a backup laser source with automatic conversion function, which can switch to a new laser source within seconds if the first laser source is damaged. This backup feature, together with the automatic calibration system, ensures minimal downtime (generally<2 hours per year). All of our laser Raman spectrometers come with a 12-month warranty, which includes the laser diode light source. The digital output of DCS is compatible with standard MODBUS or OPC protocols using RS 232, RS 485, Or Ethernet connection.

Main product functions
  1. Fixed frequency, narrowband diode laser
  2. Visual Rapid Spectral Analyzer
  3. PROSpectTM Featured Software
  4. Real-time chemometrics
  5. Code tracking
  6. Mode conversion function
  7. Multi channel function
application area
  • Refinery
  • Polymer technology
  • pharmacy
  • Food processing
  • national defense
  • Environmental monitoring
main features
  • Few or no requirements for sample conditions
  • Automatic multi-point neon lamp wavelength calibration
  • Wide standardized white light spectrum ensures stoichiometric form transfer
  • Wavelength tracking for each spectrum
  • System stability:~ 0.010 cm-1 per month
  • Real time chemometric analysis for monitoring multiple parameters
  • GRAMS/AI7TM PLSplus/IQTM compatible
  • 18 multi-channel channels
  • Ultra low maintenance workload
  • Digital or 4-20 mA analog output
  • Application Report of Online Raman Spectroscopy Analyzer
PI-200-I Online System Technical Specifications
Spectral camera aperture f/2.0
Excitation accuracy of 785nm laser light source ~4cm-1
CCD camera AndorTMtechnology
Spectral Range 250 to 1900cm-1(@1000 gr/mm)
Cooling method of camera TE refrigeration:- 55 ° C to -80 ° C
Fiber optic coupling 50 µ m x 3.3mm rear coupling
Laser (power>400 MW) 785 nanometer wavelength standard, other wavelengths are also available
temperature control Laser and Spectral Analyzer
Raman probe InPhotonics Raman Probe
Raleigh filter > 10-8
packing Industrial computer, 19ncabinet, Windows xp
Multi channel technology 18 channel capacity
Computer requirements WindowsTMXP, PC or laptop
Benchtop size configuration 50.8cm × 40.6cm × 20.3cm
Weight Benchtop configuration ~66 pounds (30 kilograms)
Power demand 85 to 265 VAC, 47 to 63 Hz, 4 amperes
CE certification adopt