summary
The MODEL 9830 iron (Fe) ion water quality online automatic monitoring instrument is independently developed by our company. The instrument uses spectrophotometry to measure the concentration of iron ions in water. This instrument is capable of long-term unmanned automatic monitoring of iron ions in various water bodies. Its measurement method and device can monitor lower concentrations of iron ions.
applicationrange
Petrochemical industry, mineral processing industry, chemical industry, metallurgical industry, electroplating industry, wastewater treatment industry, environmental protection related industries.
Instrument characteristics
•
This instrument adopts analytical methods, has strong anti-interference ability, and is not affected by color.
•
Abnormal alarms and power outages will not result in data loss.
•
After abnormal reset and power failure, the instrument will automatically eliminate residual reactants and restore working status.
•
Equipped with automatic range switching function, no manual setting required.
•
It has a reverse control function, and all functions of the instrument can be controlled through remote central control, increasing the convenience of operation.
•
Sampling methods such as timing, interval, and telemetry can be set.
•
usePanasonicImportPLCWaiting for control components reduces environmental interference and equipment failures.
•
Adopting sequential injection analysis method, the flow path is simple, the reagent consumption is low, and the degree of automation is high.
•
Adopting independently developed multi-channel valves, it has high anti-corrosion performance, long service life, and is safe and reliable.
•
Powerful data storage function, capable of storing data up to30000Group year or above.
•
Convenient data export function,UExport historical data reports directly from the disk.
Technical Parameter
•
Measurement range:0~1000mg/L
Fe2+
•Accuracy:±10%
(Indication error)
•Repeatability:≤
5%
•Measurement cycle: minimum measurement cycle
35min
•Measurement interval: Interval mode(1~9999)minExternal control mode,Hourly mode(24Hours can be set arbitrarily
)
•Correction method: manual/
Automatic mode
•Service interface:RS232perhaps
RS485
• apparent
Display: Color touch screen
•
Operation menu: Chinese
•Data Storage30000:
strip
•4Analog output: (20~)
mA
•MTBF:≥720h/
order
•Power Supply: (220±22)V AC,
(50±1)Hz
•Size: