MX-3000 chemiluminescence nitrogen analyzer
Domestic nitrogen integrated analysis instruments are widely used for detecting nitrogen content in liquid, solid, or gas samples. Compared with similar instruments at home and abroad, it hasCharacteristics such as stable and reliable performance, high analysis accuracy, and good repeatability
The total nitrogen content in the sample is rapidly measured by chemiluminescence method. The application of high-precision data acquisition and computer technology for data collection
Collection, control, and processing provide reliable assurance.
The instrument system adopts chemiluminescence method to determine the total nitrogen content, which improves the ability of Kang impurities to interfere and avoids the cumbersome operation of the titration cell by the electric quantity method and the potential impact
The instability caused by this greatly improves the sensitivity of the instrument. The key components of the system use Jin port devices, which provide reliable overall performance
Guarantee.
Instrument execution standard:
SH/T 0657-1998 Determination of Trace Nitrogen in Liquid Petroleum Hydrocarbons (Oxidative Combustion and Chemiluminescence Method)
ASTM D4629-1996
Technical parameters:
Basic parameters: Sample types: liquid, solid, and gas
Measurement method: chemiluminescence method
Sample injection volume: Solid sample: 1-20mg
Liquid sample: 5-20 μ L
Gas sample: 1-5mL
Measurement range: 0.1-10000mg/L
Measurement accuracy:
Concentration value (ppm) |
Injection volume (μ L) |
RSD(%) |
0.1 |
20 |
25 |
5 |
10 |
10 |
50 |
10 |
5 |
100 |
10 |
3 |
5000 |
10 |
3 |
Temperature control range: Room temperature~1050 ℃
Temperature control accuracy: ± 3 ℃
Gas source requirements: high-purity argon gas: purity above 99.995%,
High purity oxygen: purity above 99.995%
Power supply: AC220V ± 22V, 50Hz ± 0.5Hz, 1500 W
Dimensions: Host: 305 (W) × 460 (D) × 440 (H) mm
Temperature control: 550 (W) × 460 (D) × 440 (H) mm
Weight: Host: 20kg
Temperature control: 40kg
Requirements for instrument completeness: host (data acquisition, control system), cracking furnace, automatic sampler (gas, liquid), computer (including operating software)
Wait.
Other optional components: solid sampler, gas sampler
The instrument is suitable for measuring the total nitrogen of paraffin oil, diesel, gasoline, lubricating oil, fuel oil, liquefied gas, as well as other oil products, chemical raw materials, and finished products
Content.