Zhengzhou Chuangyu Technology Co., Ltd
Home>Products>Methane/hydrocarbons/transformer oil gas analysis chromatograph
Methane/hydrocarbons/transformer oil gas analysis chromatograph
The 2060F methane/hydrocarbon/transformer oil gas analysis chromatograph instrument adopts a microcomputer automatic control system and a Chinese LCD
Product details

Methane/hydrocarbons/transformer oil gas analysis chromatography

Gas analysis chromatography can be used for underground gas analysis in coal mines, and for the analysis of dissolved gases in transformer oil, including H2, O2, CO, CO2, CH4, C2H6, C2H4, and C2H2. The analysis of H2, O2, N2, CO, CO2, and hydrocarbon gases in gases such as coke oven gas, petroleum gas, and liquefied gas is widely used in various analyses of industrial gases.
The instrument adopts a microcomputer automatic control system and a large Chinese LCD screen display. The design is flexible and reasonable, and it is equipped with thermal conductivity and dual hydrogen flame detectors, methane conversion furnace, gas injection valve, and various chromatographic columns. The gas path process used achieves a single injection and three detector technology to complete the analysis of constant and trace components of gases such as H2, O2, N2, CO, CO2, CH4, alkanes, alkenes, alkynes, etc.
Methane/hydrocarbons/transformer oil gas analysis chromatograph

Main performance characteristics
The gas path process of ※ * can flexibly adjust the separation degree of hydrocarbon components and CO, CO2, with accurate quantification and high sensitivity;
Using methane conversion equipment and triple detector technology to detect trace amounts of CO and CO2;
The detectors are designed with a modular manufacturing process, featuring high sensitivity, low noise, and a wide linear range. Different detectors can be configured according to analysis needs;
※ Automatic/manual injection function, can be freely selected;
The analysis cycle is within ten minutes, and the analysis speed is fast;
※ Extended functions: equipped with electronic capture detector to determine tracer gas SF6, flame photometric detector to measure gases such as H2S, SO2, etc;
Methane/hydrocarbons/transformer oil gas analysis chromatograph

Technical Specifications
※ Detection concentration: H2 ≤ 5ppm CO, CO2 ≤ 2ppm hydrocarbons ≤ 0.1ppm
Specification
※ Temperature control indicators
Temperature control area: chromatography column box, thermal conductivity, hydrogen flame, conversion furnace
Temperature control range: 6 ℃ to 399 ℃ at room temperature (increment 1)
Temperature control accuracy: better than ± 0.1 ℃
※ Detector indicators
Sensitivity of thermal conductivity cell S ≥ 5000mv. ml/mg (n-hexadecane)
Hydrogen flame detection limit Mt ≤ 1 × 10-11g/s (C6H6)
Gas analyzer configuration: TCD+double packed column+single six way valve injection