Micro moisture analyzer
Brief description: The trace water analyzer uses the principle of Coulombic titration to determine the trace water content in samples. Compared with s
Product details
Overview of Trace Moisture Analyzer
The TW-2026WS micro water analyzer uses the principle of coulometric titration to determine the trace water content in samples. Compared with similar excellent products at home and abroad, it has the advantages of high sensitivity, fast electrolysis speed, short equilibrium time, accurate blank deduction, and accurate and reliable analysis results. It is a high-quality analytical instrument for departments such as petroleum, chemical, power, environmental protection, medicine, and scientific research.
The instrument adopts a brand new design concept and operates on a full touch screen. The instrument has a small size and a high degree of circuit integration. The electrolytic cell is designed to prevent moisture from entering, greatly eliminating the impact of high humidity environments on the analysis process. Powerful software with blank deduction function ensures accurate deduction of blanks within 10 minutes of sample enrichment time. The equilibrium point can also be manually or automatically adjusted according to different sensitivity reagents to facilitate the analysis process.
Micro moisture analyzerTechnical indicators
1) Display mode: High definition color touch screen display with dynamic display of electrolytic curve
2) Titration method: Electric titration (Coulombic analysis)
3) Measurement range: 1ug~100mg (typical value 10ug~100ug)
4) Resolution: 0.1ug
5) Accuracy: 3ugH2O~1mgH2O ± 2ug,
0.3% above 1mg H2O (excluding injection error and environmental humidity error)
6) Electrolytic current: (0~400)mA , Automatic tracking control.
7) Sample type: Solid, Liquid, Gas
8) Data storage: 1000 test records
9) Status indication: dynamic curve, text display
10) Mixing speed regulation: sliding touch panel speed regulation
11) Date and Time: Ten Years of Power Failure, Real Time of Normal Operation
12) Printer: Microprinter
13) Environmental temperature: (5-40) ℃
14) Environmental humidity: ≤ 85%
15) Power supply: AC220V ± 10% 50Hz ± 5%
16) Power: 40VA
17) Dimensions: 35 × 27 × 25 (length × width × height)
Online inquiry