1. Working conditions
Environmental temperature: 5-35 ℃ Relative humidity:< 85%
Power supply: 220V±10% 50Hz ± 1Hz
No significant vibration
There is no strong magnetic field interference except for the geomagnetic field.
2. Main technical indicators
Measurement range: Conductivity: (0~199.9)us/cm ; (200~1999)us/cm ;
(2.00~19.99)ms/cm ; (20.0~199.9)ms/cm ;
Resolution: 0.1/1us/cm ; 0.01/0.1ms/cm ;
Accuracy: Electrometer: ± 1.0% FS; Supporting facilities: ± 1.5% FS
Temperature compensation range: 0-50 ℃ (manual)
Temperature coefficient: 2.0%/℃
Optional electrode specification constant: 0.1, 1, 10
Data storage: 50 sets
Storage content: number, measurement value, measurement unit, and temperature value
Power supply: DC9V adapter
Dimensions and weight: 220 * 200 * 80mm (length * width * height) 0.7 kg