Shanghai Tongli Instrument Co., Ltd
Home>Products>Mixier's Liquidew I.S. liquid moisture analyzer
Product Groups
Firm Information
  • Transaction Level
    VIP member
  • Contact
  • Phone
  • Address
    Building 9, Lane 6558, Tingwei Road, Jinshan District, Shanghai
Contact Now
Mixier's Liquidew I.S. liquid moisture analyzer
Cantilever beam sensor
Product details


Mixier's Liquidew I.S. liquid moisture analyzer can measure the moisture content in liquids online. It can continuously measure various non-polar liquids online, including flammable liquids and hazardous applications, petrochemical industry, power plants, and pharmaceutical industry.







Highlights

  • Simple and efficient operation with minimal maintenance
  • Suitable for renovation projects, easy to integrate into existing sampling or branch systems
  • The measurement range of water vapor in liquid is from 0.01 ppmW to saturation point
  • Multi channel, supporting up to four completely independent measurement channels
  • Durable and sturdy design, capable of measuring large quantities of non-polar liquids without interference over time
  • Two 4-20 mA adjustable outputs and digital Modbus RTU communication
  • Reliable replaceable sensing components obtained through Michell's verification and replacement service
  • NIST and NPL traceability verification
  • Provide EExd model
Liquidew I.S.

Liquidew I.S.

TECHNICAL INDEX

sensing technology Michell ceramic humidity sensor
Sensor version Easidew PRO I.S
range Please specify the range of 0.001~1000 ppMW, as the actual range depends on the water vapor capacity of the sampled liquid
Verify range -100 ~ +20°Cdp
accuracy the dew point: ± 1 ° C between -60 ° C and+20 ° C Cdp
Moisture content: ± 10% of reading
the dew point: ± 2 ° C between -60.1 and -100 ° Cdp
resolution 0.1 ° C between+20 ° C and -100 ° C Cdp
temperature measurement Pt100
temperature range -20 ~ +70°C
operation temperature -20 ~ +60°C
Sample flow rate Minimum 0.01 l/min, maximum 10 l/min
0.1~0.3 l/min recommended
check Tracing back to humidity standards in the UK (NPL) and the US (NIST)
I. S. Certificate ATEX certification authorized by EECS can be used for hazardous environments Ex II 1G EEx ia IIC T4
CSA certification IS Class I, Division 1, Groups A, B, C & D, T4
IECEx: Ex ia IIC T4
FM: IS/I/1/ABCD/T4 Ta=+60°C - Ex90385; Entity; IP66
Controller
display Two rows of 6-digit LEDs display moisture content/dew point (user selected) and temperature
Analog Output Two channels of 4-20 mA (maximum load 500 Ω) - user set parameters, units, and ranges
Digital Output RS485 Modbus RTU
display mode Moisture content (ppMW)
Dew point (° C or ° F)
Temperature (° C or ° F)
display resolution 0.1°Cdp, 0.1°Fdp, 0.001 ppmw, 0.1 ° C temperature
call the police Four alarm relays Trigger type and set point selectable
Two way mode C contact level 10A, 240 V AC or 8 A, 24 V DC non inductive load
Two way mode A contact level 5A, 240 V AC or 4 A, 24 V DC non inductive load
I. S. Safety Barrier Isolation type, with built-in control module
power supply 85-265 V AC 50/60Hz or 10-72 V DC maximum power consumption 10 W
casing 19 "rack control panel
Size 132 x 483 x 256mm (h x w x d) (with a minimum clearance of 100mm at the rear end)
work environment Indoor, safe area, 0 ~ +50°C, < 90% RH
A comprehensive sampling system
casing 304 stainless steel (EN1.4301) cabinet
316 stainless steel (EN1.4401) cabinet is optional; Made entirely of stainless steel material; Built in installation panel electroplating treatment; Indoor installation can choose open panel type
Size 800 x 600 x 300mm (h x w x d)
install Stainless steel wall mounted installation bracket
protection grade IP66
temperature control Heater/temperature regulator option, can be a fixed control temperature point of+20 ° C or an adjustable temperature range of 0~50 ° C
Heater power supply 110/120 or 220/240/255 V AC, 50/60 Hz power supply with a maximum power consumption of 100 W
working condition Shaded areas, on land or at sea- When the ambient temperature is above+50 ° C (-40~+60 ° C max. transient), it is recommended to use the refrigeration option
Liquid cooling option Process liquid temperature>+40 ° C, it is recommended to cool down first
Factory water supply, water temperature ≤+30 ° C