OXITESTThe analyzer accelerates the oxidation process of the sample by controlling two factors, temperature and pressure. At a constant temperature, the absolute pressure of oxygen in the titanium reaction chamber is measured to reflect the absorption of oxygen by the sample. And automatically generate induction period values (IP) through software.
OXITESTThe analysis method of the analyzer isAOCSOfficial methodCd 12c-16.
main features
No need for sample preparation, measure the oxidation stability of the undisturbed sample
The oxidation chamber, sample tray, and reaction chamber cover are made of titanium metal
OXISoft™ The software is intuitive and easy to use, with comprehensive functions
8Synchronous measurement of samples: Each instrument has 2 independent reaction chambers, and one computer controls 4 instruments
application area
Food, feed, beverages, environmental protection, pharmaceuticals and chemicals, etc
Measure induction period IP
The oxidation curve of the sample: The induction period is the time required for the sample to reach the oxidation starting point.
The sudden change in detectable rancidity or oxidation rate indicates the end of the induction period.
The longer the induction period, the stronger the antioxidant capacity of the sample.
Technical Parameter
Number of reaction chambers |
2 |
Single lens reflex chamber capacity |
100ml |
Pressure Range |
0~8bar |
temperature range |
Room temperature to 120 ℃ |
Set maximum temperature deviation |
≤0.5℃ |
Reproducibility of Temperature Setting |
≤±0.2℃ |
interface |
USB |
connectivity |
LANOr Wi Fi access to the cloud |
power |
900W |
voltage |
115V/60Hz – 230V/50-60Hz |
weight |
16.5kg |
size |
365x190x485 mm |
Overvoltage protection |
safety valve |
High temperature protection |
Visual alarm |
sensor fault |
Visual alarm |
Origin and Manufacturer: VELP, Italy