Brief description: The total sulfur analyzer for oil products uses ultraviolet fluorescence method to determine the total sulfur content, which improv
Product details

Oil Total Sulfur Analyzer

Oil Total Sulfur AnalyzerSH/T 0689-2000 Determination of Total Sulfur Content in Light Hydrocarbons, Engine Fuels, and Other Oil Products (Ultraviolet Fluorescence Method)
ASTM D 5453-2006 Standard Test Method for Total Sulfur in Light Hydrocarbons, Spark Ignition Engine Fuel, Diesel Engine Oil, and Engine Oil by Ultraviolet Fluorescence Measurement; European standard number EN ISO 20846:2004 title Petroleum products Determination of sulfur content in motor vehicle fuels Ultraviolet fluorescence analysis method; GB 17930-2011 Automotive Gasoline; GB 19147-2009 Automotive Diesel Fuel; ISO 20846-2011 Petroleum products - Determination of sulfur content in automotive fuels; BS EN 15486-2007 Ethanol is a mixed component of petroleum Determination of sulfur content, GB/T 11141-2014 Determination of trace sulfur in light olefins for industrial use

System configuration:
Standard configuration: printer+computer+TS-3000 system+liquid sampler
Other optional components: solid sampler, gas sampler
Easy to operate:
Based on the Chinese operating interface of Windows (7, XP, Me, 98), it makes the operation more convenient and fast. You only need to tap the mouse to complete all parameter settings and condition selections, and the computer controls the collection, processing, storage, and printing of data.

Technical Parameter
Basic parameters: Sample types: liquid, solid, and gas
Measurement method: UV fluorescence method (S)
Sample injection volume: Solid sample: 1-90mg
Liquid sample: 5-20 μ L
Gas sample: 1-50mL
Measurement range: 0.1-50000mg/L (high concentration can be diluted, low concentration gas samples can reach 0.1ppm)
Measurement accuracy:

Concentration value (ppm)

Injection volume (μ L)

RSD(%)

0.2

20

50

5

10

10

50

10

5

100

10

3

5000

10

3

Temperature control range: room temperature~1300 ℃

Temperature control accuracy: ± 1 ℃

Gas source requirement: argon gas: purity above 99.9%
Oxygen: purity above 99.9%

Power supply: AC220V ± 22V, 50Hz ± 0.5Hz, 1500 W

Dimensions: Host: 305 (W) × 460 (D) × 400 (H) mm
Temperature control: 550 (W) × 460 (D) × 400 (H) mm

Weight: Host: 20kg
Temperature control: 40kg