The PPS series programmable AC source meter is a high-quality AC programmable voltage regulator programmable power supply, featuring intelligent programmability, integrated source meter, high stability, high purity, high measurement accuracy, remote voltage sampling, and ultra-thin chassis design. This power supply provides pure AC power for various types of loads and can simulate continuous changes in grid voltage to achieve power characteristic testing of loads. Especially suitable for high demand applications such as metrological verification, scientific research experiments, and precision testing.
Main features:
The output voltage can be set once or continuously changed through programmable control to meet the power characteristic testing requirements of various loads;
The start/end phase can be set between 0 ° and 359 ° to simulate sudden waves or notches;
Low distortion, high purity, suitable for harmonic testing occasions;
◆ Rich measurement functions: In addition to basic parameters such as voltage, current, power, power factor, and frequency, it can also measure parameters such as voltage/current peak value, voltage/current peak ratio, and total harmonic distortion of voltage/current, truly integrating the source and meter;
Voltage remote sampling function, avoiding line loss and meeting the needs of high-precision testing occasions;
◆ Equipped with remote input (TEST, RESET, call memory group 1-50) and remote output (PASS, FAIL, TEST IN PROCESS) functions, convenient for online automation control.
|
model |
PPS1005 |
PPS1010 |
||
|
output capacity |
500VA |
1KVA |
||
|
each other |
1φ |
|||
|
Output voltage |
2V~150V,2V~300V |
|||
|
Maximum current |
0-150V |
4.6A |
9.2A |
|
|
0-300V |
2.3A |
4.6A |
||
|
OUTPUT FREQUENCY |
Range Range |
40.00HZ~500.0HZ |
||
|
Stability |
≤ 0.03% reading+1 digit |
|||
|
Total harmonic distortion |
40Hz~100Hz |
≤0.5% |
||
|
101Hz~500Hz |
≤1% |
|||
|
voltage stability |
≤0.2% |
|||
|
Transient response time |
<400μs |
|||
|
Power Efficiency |
≥80% |
|||
|
communication |
RS232/485 |
|||
|
Test project |
V、A、W、PF、Hz、Ap、Vp、Acf、Vcf、Athd、Vthd |
|||
|
voltmeter |
range |
75V/150V/300V |
||
|
resolution |
0.1V |
|||
|
Basic accuracy |
0.4% reading+0.1% range+1 word |
|||
|
ammeter |
range |
0.5A/A/10A |
||
|
resolution |
0.001A(0.001A~9.999A); 0.01A(10.00A~19.99A) |
|||
|
Basic accuracy |
0.4% reading+0.1% range+1 word |
|||
|
power meter |
resolution |
0.1W(0.1W~999.9W); 1W(1000W~1999W) |
||
|
Basic accuracy |
± [(0.001/reading)+0.001 × (4+0.02f)+1 word] |
|||