summary
PTC-30A typeChromatographic pre-treatment deviceSolid liquid integrated blowing and capturing instrumentIt is a 76 bit device with an electronic cold trapSolid liquid integrated blowing and capturing instrumentSupports automatic and precise sampling of liquid samples into blowing bottles for blowing and capturing, as well as automatic addition of distilled water for shaking and mixing of solid samples for direct blowing and capturing.
Performance indicators:
PTC-30A typeChromatographic pre-treatment deviceSolid liquid integrated blowing and capturing instrument
1. Sample grade: 76th grade, 40ml VOA sample bottle
2. Purging tube temperature: room temperature~100 ℃, step size 1 ℃
3. Temperature control range of cold trap (electronic semiconductor refrigeration): -40 ℃ to room temperature, step size 1 ℃, no need for liquid nitrogen refrigeration
4. Desorption temperature of capture tube: 150-380 ℃, heating rate>3000 ℃/min; Reduce from 250 ℃ to 35 ℃ within 90 seconds
5. Heating temperature of pipeline: room temperature~220 ℃, step size 1 ℃
6. Heating temperature of water purifier: room temperature~400 ℃
7. Cleaning distilled water temperature: room temperature~90 ℃, step size 1 ℃
8. Solid sample bottle heating: room temperature~100 ℃, step size 1 ℃
9. Solid sample mixing: 5ml distilled water, shake and mix
10. Glass blowing tube: 5ml blowing tube
11. Catching pipe size: stainless steel material Φ 3mm × 180mm × 0.1mm
12. Purging flow rate: 10-150ml/min
13. Timing range: 1 second to 100 minutes
14. Timing error: better than 0.1%
15. Air circuit pressure resistance:> 0.4MPa
16. Synchronous signal output: 2 switch signals, with a linkage switch time of 1-2 seconds
17. Power supply: AC220V 50Hz
18. Power:< 800W
19. Instrument size: 640mm * 750mm * 510mm
20. Instrument weight: approximately 42Kg
Environmental adaptability
1) Backblowing carrier gas pressure: ≤ 0.4MPa
2) Environmental temperature: 5-35 ℃
3) Relative humidity: ≤ 85%
Main functions and features
1. Strong universal performance: can be used in combination with any brand of gas chromatograph (GC) and (GC-MS)
2. Easy to operate and highly automated: simply place the sample bottle in the sample tray, and other operating software will complete it, so the experiment has good repeatability; The host is equipped with a large size
Touch LCD screen, rich interface information, fully tracking temperature and operation commands throughout the process
3. The cold trap adopts semiconductor cooling and air cooling, with a low cooling temperature of up to -40 ℃ (at room temperature of 20 ℃)
4. The heating of the capture tube is achieved through direct resistance heating, with a heating rate greater than 3000 ℃/min
5. The built-in foam sensor can detect the foam in the purging pipe, prevent the way of polluting the sample, and protect the entire analysis system
6. Users can customize the cleaning frequency of the sampling needle and blowing tube
7. Solid sample bottles can be heated and directly injected with distilled water for oscillation, mixing, and blowing
8. The gas separation between the capture tube and the blowback tube reduces cross contamination between samples. In the blowback cycle, compounds blown out of the capture trap will not flow into the blowback
Guanzhong
9. The water trap removes water vapor at the blowing end, greatly reducing the impact of water vapor on GC and GC/MS
10. The sample tubing is made of PEEK material and treated with silane inert stainless steel to reduce residual pollution
11. All pipelines and six way valves are temperature controlled for heating, eliminating system cold spots and reducing sample loss
12. Provide a synchronization interface that can start both the chromatography and workstation simultaneously during sample injection
Common application areas
By combining with GC or GC/MS, it can be widely applied in environmental analysis, such as the analysis of organic pollutants in drinking water, wastewater, and soil, as well as the analysis of volatile compounds (such as odor components) in food.
fully automaticSolid liquid integrated blowing and capturing instrument
List of applicable main standards
GBT5750.8-2006 Standard Test Method for Drinking Water, Determination of Organic Matter Index
HJ605-2011 Determination of volatile organic compounds in soil and sediment by blow sweep capture/gas chromatography-mass spectrometry method
HJ639-2012 Determination of volatile organic compounds in water quality by blow sweep capture/gas chromatography-mass spectrometry method
HJ686-2014 Determination of volatile organic compounds in water quality by blow sweep capture/gas chromatography-mass spectrometry method
HJ735-2015 Determination of volatile halogenated hydrocarbons in soil and sediment by blow sweep capture/gas chromatography-mass spectrometry method
HJ788-2016 Determination of Acetonitrile in Water Quality by Sweep Capture/Gas Chromatography Method.