1. Product Name | |||
1.1 Name 1.2 Model |
Three box cold and hot shock test chamber (water-cooled) PW-CTS3-80-40 |
||
2. Purpose and Description | |||
2.1 Scope of Application |
1. This equipment is suitable for simulating the adaptability of various electrical and electronic products, as well as other products, components, and materials to rapid changes in ambient temperature for reliability testing. In some cases, it can also be used for environmental stress testingfilterExperiment. 2. Other uses may result in personal injury and equipment damage! |
||
2.2 Limitations on Test Samples |
1. Testing and storage of samples of flammable, explosive, and volatile substances 2. Testing and Storage of Corrosive Substance Samples Testing or Storage of Biological Samples 3. Testing and storage of strong electromagnetic emission source samples |
||
3. Capacity and size | |||
3.1 Volumetric capacity |
make an appointment80L |
||
3.2 Inner box dimensions |
W (width) 400 mm * H (height) 500 mm * D (depth) 400 mm |
||
3.3 Outer box dimensions |
W (width) 1450mm * H (height) 1950mm * D (depth) 1750mm reminder:Please refer to the final design confirmation of the three views for external dimensions! |
||
4. Performance (the following indicators are for no-load performance) | |||
4.1 Condition requirements |
ambient temperature10℃——28℃,When unloaded (high ambient temperature affects cooling effect) |
||
4.2 Structural Types |
Three boxes (high-temperature storage area, low-temperature storage area, sample testing area) static airflow impactThe sample does not move during the high and low temperature conversion process. |
||
4.2 Temperature Range |
High temperature studio: heat storage rangeFrom 60 ℃ to+180℃ Low temperature studio: cold storage range0 ℃ to -55 ℃ Sample testing area: high temperature impact range60 ℃ to 150 ℃, low temperature impact range 0 ℃ to -50 ℃ |
||
4.4 Temperature fluctuation |
Temperature fluctuations:±0.5℃ |
||
4.5 Temperature deviation |
Temperature:≤± 1 ℃ (the average temperature displayed by the studio temperature controller minus the average temperature measured at the center point) |
||
4.6 Temperature uniformity |
≤ 2 ℃ (temperature uniformity is the arithmetic mean of the difference between the highest and lowest temperatures measured in each test) |
||
4.7 Heating and cooling capacity of high and low temperature storage boxes |
High temperature chamber: The heating rate is determined by+From 20 ℃ to+180 ℃ for approximately 45 minutes, with an average nonlinearity throughout the entire process Low temperature box: cooling rate is determined by+20 ℃~-65 ℃ for about 60 minutes, with an average nonlinearity throughout the entire process |
||
4.8 Test box indicators |
Impact conversion time:Adjustable from 2 seconds to 3 minutes (defined as the high and low temperature conversion time of the test sample in 1.6.5 of GB2423.22, including the time to return to room temperature) Temperature recovery time: ≤ 3 minutes (0.1 * sample test time) (empty load), (according to1.6.4 in GB2423.22 is defined as the conversion between high and low temperaturesTime for the temperature of the test chamber to reach the set value Sample test lower limit time: greater than or equal to30 minutes (holding time after the temperature of the test chamber reaches the set value) Continuous working time of the test chamber: greater than500 hours (the time during the experiment when the temperature will not be affected by the defrosting of the evaporator and the recovery of high and low temperature conversion) |
||
4.9 Noise |
noise≤ 70dB (measured in a soundproof room with low echo at an ambient temperature of 25 ℃; average value of 8 points tested; each test point is 1 meter horizontally from the noise source and 1 meter above the ground). Noise reduction scheme: 1. The main source of noise, such as the compressor equipped with damping springs, can reduce noise and minimize the destructive factors caused by vibration, thereby extending the service life of the equipment. 2. High quality thickened sound-absorbing cotton is pasted on the inner wall of the refrigeration unit, which can effectively reduce the noise effect of the refrigeration unit. 3. Vibration isolation hoses are standard for connecting compressors and pipelines. |
||
4.10 Power consumption |
The refrigeration system of this machine is controlled by cold balance technology, based on the temperature requirements inside the boxPID control solenoid valve switch controls the flow of refrigerant to achieve the effect of cooling without heating and heating without coolingThe power consumption is significantly reduced compared to traditional thermal balance (where the cooling capacity is fixed and balanced by heating and humidification). |
||
4.12 Meet the test standards |
GB2423.22 Temperature Change Test GB2423.1 Low temperature test method GB2423.2 High temperature test method GJB150.5 Temperature Shock Test |
||
5. Structural part |
notes |
||
5.1 Structural characteristics |
This device is designed with functional modules, and the main circuit and refrigeration configuration system are placed independently in functional sections (with the circuit part having independent space for dust prevention and heat dissipation), achieving isolation between the circuit and refrigeration system, making the system safer and more reliable |
||
5.2 Structural Materials and Processes |
Outer box: The frame is made of profiles, and the overall structure of the machine components is assembled. The cold plate is coated with electrostatic powder paint, and the color is computer white, which is neat, beautiful, and stable. Inner box:SUS304 stainless steel plate seamless welding forming The circuit and refrigeration cover plate adopts a key switch design, which is more convenient for maintenance and upkeep |
|
|
5.3 Thermal insulation enclosure structure |
High temperature resistant insulation material: rigid polyurethane foam+Fiberglass |
||
5.4 Air conditioning channel |
The high temperature zone, low temperature zone, testing zone, air damper, and air circulation system are shown in the following figure
|
||
5.6 Door |
Left single door opening, door frame equipped with anti condensation electric heating device, using double sealed foam silicone sealing around to ensure separation between indoor and outdoor areas |
|
|
5.7 Control Panel |
Embedded built-in touch screen for better screen protection |
|
|
5.8 Casters |
Fuma wheels |
|
|
5.9 Door handle |
|
||
5.11 Test hole |
This machine is on the left50MM1 pieces. |
|
|
5.12 Product Placement Rack |
bearing:20kg |
|
|
5.13 Air source intake structure |
Filter pressure reducing device |
|
|
5.14 Drainage |
Condensate drain outlet |
|
|
6. Circuit system and main components |
notes |
||
6.1Distribution process |
according toGB7251, GB2681, GB/T2682, GB50171, GB50256 are combined with the actual situation of the company and equipment to ensure that the electrical equipment meets the design control requirements for primary and secondary installation and wiring, and is neat, beautiful, and easy to inspect. |
|
|
6.2power switch |
|
||
6.3 Weak current power supply circuit |
switch power supply |
|
|
6.4 Load Circuit |
AC contactor |
|
|
6.5 Sensor |
T-type thermocouple |
|
|
6.6 Temperature control circuit |
Solid state relay(SSR) |
|
|
6.7 Safety protection circuit |
Air Switch |
|
|
7. Supply air temperature measurement system and main components |
notes |
||
7.1 Air supply method |
centrifugal fan |
||
7.2 Cycle Motor |
Low voltage asynchronous high-temperature resistant long shaft motor |
|
|
7.3 Wind turbine |
Multi blade centrifugal circulating fanMade of aluminum alloy with reinforced axis and high and low temperature resistant rotating blades |
|
|
8. Freezing system and main components |
notes |
||
8.1 Refrigeration method |
This system adopts mechanical compression refrigeration, consisting of a compressor, condenser, expansion valve, and evaporator. The refrigerant boils at pressure and temperature, lower than the temperature of the object or fluid being cooled. The compressor continuously draws in the vapor generated in the evaporator, compresses it to the condensing pressure, and then sends it to the condenser for isobaric cooling and condensation into liquid under pressure. The heat released during refrigerant cooling and condensation is transferred to the cooling medium(usually water or air), the condensation temperature corresponding to the condensation pressure must be higher than the temperature of the cooling medium. The condensed liquid enters the evaporator through an expansion valve or other throttling components |
The cooling medium of this device is water (water-cooled), using cascade refrigeration |
|
8.3 System Process |
1) Oxygen free copper tube gas shielded welding: This method avoids the damage to the refrigeration system and compressor caused by the generation of oxides on the inner wall of copper pipes through traditional welding methods. 2) Vibration reduction measures: A) Compressor: Spring damping, thickened and reinforced chassis to avoid chassis resonance B) Refrigeration system: Optimize piping paths to reduce bends. 3) Insulation: Using imported Folex CLASS0 grade high-quality insulation pipes to prevent condensation and reduce losses 4) Control of key parameters in the system Compressor return and exhaust temperature control: equipped with independent return and exhaust temperature control circuits, the controller intelligently controls the spray cooling solenoid valve to ensure that the return temperature does not exceed the operating range of the compressor; In harsh working conditions and abnormal situations, excessive exhaust is the main cause of compressor failure. Equipped with an exhaust temperature sensor, the compressor is protected by an alarm for high exhaust and shutdown |
|
|
8.4 Compressor |
Fully enclosed piston compressor |
|
|
8.5 Refrigerant |
Non fluorine environmental protectionHFC R404A, R23 (both with an ozone depletion index of 0) |
|
|
8.6 Condenser (air-cooled) |
Air cooled high-efficiency finned condenser. |
|
|
8.7 Evaporator |
Efficient multi-stage evaporator with hydrophilic film fins (thickened fins) |
|
|
8.8 Freezing pipeline switch |
Battery valve (Japan Road Palace) |
|
|
8.9 Throttle device |
Thermal expansion valve |
|
|
Hot gas bypass valve(PTV10W) |
|||
8.10 Heat exchanger |
plate heat exchanger |
|
|
8.11 Other attachments |
All are high-quality products from top international brands(The image on the right shows a dry filter) |
|
|
oil separator |
|
||
9. Heating and pneumatic system and main components |
notes |
||
9.1 Heating system |
electrical heating Temperature control method: ControllerPID uses solid-state relays (contactless periodic pulse switches) to control the heat generation of the heating tube and stabilize the temperature |
||
9.2 Heating tube |
nichrome wire |
|
|
9.3 Air door cylinder |
Adopting high-performance cylinders from Yadeke |
|
|
10. Controller |
notes |
||
10.1 Control mode |
High precision resistive touch screen+UNIQUE temperature control module from Japan. |
|
|
10.2 Display |
7-inch TFT true color LCD display screen (Resolution 800 * 480) |
||
10.3 Power input |
3 sets of inputs, supporting dry wet bulb PT100, thermocouple T, K |
||
10.4 Accuracy |
0.1 high-precision sampling |
||
10.7 Output Mode |
PID+SSR/SCR automatic bidirectional synchronous output |
||
10.8 Software Features |
Programming control software system1 set, with the following main functions: 1. Temperature and humidity curve: Can easily generate temperature and humidity curve files automatically and save them asXLS table; 2. Comprehensive interface: causeThe USB port allows for easy export of historical curves and other data using a USB flash drive. Stay outsideLAN standard network interface: No need to configure a server specifically, it can be easily connected to the company's computer local area network. One can connect and monitor up to 16 devices, which is convenient and fast. be equipped withRS-485 or RS232 standard communication interface, capable of online control and management with computers 3. Real time monitoring can achieve simultaneous monitoring of the operating status of 1-16 controllers, such as real-time data of monitoring controllers, signal point status, actual output status, etc 4. Timing control Two sets of timing output control interfaces, paired with 10 time control modes, can provide switch signal output according to program requirements. Note that this is only a control signal, and the test power supply inside the box needs to be separately connected and equipped with a dedicated power supply system 5. Fault display 16 sets of fault alarm outputs, providing user-friendly prompts in both Chinese and English for the cause of the fault and troubleshooting methods |
|
|
11. Protection function |
notes |
||
11.1standard configuration |
1. Test sample over temperature protection (independent) |
|
|
2. Main power supply with leakage protection circuit breaker switch |
|||
3. Over temperature protection switch in high temperature area |
|
||
4. Compressor overload and overheating |
|
||
5. Compressor high pressure protection |
|
||
6. System overcurrent protection |
|||
7. Low temperature zone over temperature protection switch |
|||
8 Compressor Exhaust Temperature Protection |
|||
9. Compressor return air temperature protection |
|||
10 phase sequence protection |
|||
11 Fan overload protection |
|||
12. Other components |
notes |
||
12.1 Optional Accessories |
not have |
||
12.2 Shipping and Acceptance Data |
1. Operation manual and maintenance manual 2. Control system software and software documentation 3. Equipment drawings: circuit diagrams, etc 5. Acceptance form, etc |
||
13. Installation conditions for use |
notes |
||
13.1. Equipment usage environment |
1. Environmental temperature: 5℃-35℃ 2. Relative humidity: not exceeding 85% R.H 3. Atmospheric pressure: 80kPa~106kPa 4. Flat and vibration free ground; 5. Choose well ventilated areas without direct sunlight or direct radiation from other heat sources; 6. There is no strong airflow around: When the surrounding air needs to flow forcibly, the airflow should not directly blow onto the box; 7. There is no strong electromagnetic field influence around; 8. There is no high concentration of dust or corrosive substances around |
||
13.4. Compressed Air |
This device requires clean compressed air to be provided 1) Pressure: 0.5~0.7Mpa 2) Flow rate: approximately 1300 liters per minute |
The compressed air device is provided by the customer themselves |
|
13.2. Use power supply voltage |
1. Connect to a power supply of 380V AC (± 10%) Three-phase line+Protect the grounding wire, with a grounding resistance of ≤ 4 Ω; Power switch usage100A air switch 2. Maximum power of the machine:make an appointment 30 KW 3. Maximum operating current:make an appointment 40 A 4. Power frequency: 50±0.5Hz |
Attention customers, please lead the main power cable to the equipmentWithin a range of 3 meters! |
|
13.3. Placement of test products |
The product should not be placed on the same horizontal plane for more than two-thirds of the area, and the air outlet and return outlet should not be blocked (at least from the air outlet and return outlet)10CM distance) |
||
14. After sales service |
notes |
||
14.1 |
Free training for customers on operators and maintenance personnel(located in the location of our company or client) The warranty period of the equipment shall come into effect from the date of acceptance and shall be one year. During the warranty period, if the equipment malfunctions due to non-human operation, the seller shall be responsible for free repair or replacement of parts. Outside the warranty period, maintenance work on the equipment will incur a cost fee for the replacement parts and a certain service fee. |
||
15. Daily maintenance and upkeep of equipment |
notes |
||
15.1. Execution standards |
Regularly clean the dust from the condenser and compressor to prevent dust from affecting the cooling effect of the condenser. 2. Check the compressed air pressure before starting up |
Configuration List
Main Material List | |||||
name |
brand |
specifications(W*H*D) |
origin |
quantity |
notes |
AStructural part | |||||
box |
PW |
Integral |
China |
1type |
|
air handling unit |
PW |
stainless steel |
China |
1type |
|
Unit and control box |
PW |
Cold rolled steel plate baked paint |
China |
1group |
|
Double door |
PW |
Color board |
China |
1type |
|
Circulating motor |
Yutian |
1HP |
China |
3group |
|
BRefrigeration section | |||||
compressor |
Hitachi |
vortex |
Guangzhou |
2tower |
Japanese sole proprietorship |
plate heat exchanger |
Yuanzhuo |
Supporting facilities |
China |
1type |
|
condenser |
Yongqiang |
Water-cooled shell and tube type |
Taiwan |
1group |
|
evaporator |
Zhongli |
3/8Fin type |
Taiwan |
1group |
|
oil separator |
Emerson |
AWseries |
Mexico |
2type |
|
dryer |
danfoss |
DML |
Denmark |
4group |
|
solenoid valve |
Sporlan |
SEV/EVR |
Europe |
4type |
|
expansion valve |
danfoss |
TES |
Denmark |
6type |
|
Evaporator Pressure Regulator |
danfoss |
KVP |
Denmark |
1cover |
|
Refrigerant and consumables |
DuPont |
R404A R23 |
America |
2type |
|
CControl part | |||||
controller |
PW-M1TETS |
PW1000 |
China |
1tower |
|
phase sequence relay |
CHIN |
China |
1individual |
||
switch power supply |
Mingwei |
DC24V2.1A |
Taiwan |
1type |
|
sensor |
OMi Qie |
T |
Switzerland |
1type |
|
power switch |
DELIXI |
3P 100A |
China |
1type |
|
Contactor |
literati |
SP25 |
France |
6type |
|
Heating pipe |
Longxing |
Finned nickel chromium alloy |
Taiwan |
6branch |
|
Solid state relay |
Xinling |
SSR |
China |
3group |
|
cable |
Golden Dragon Feather |
0.75~12square |
China |
1cover |
characteristic:
1. High environmental requirements:5~30℃
2. By controlling the refrigerant output of the compressor to balance the temperature, it effectively saves the loss of heat balance temperature;
3. The three box impact box directly impacts the high and low temperature, reducing the rapid frosting of the evaporator caused by the transition to normal temperature and reducing testing time. It can achieveDefrost once every 200 times;
4 Scalablelarge screen10 inch, independently developed control operating system with strong user experience, and lifetime free upgrade service;
5 The low power of the refrigeration compressor reduces the energy consumption of the entire machine;
6 Noise: Whole machine65dB;