Product Overview
The Picarro G2207-i gas concentration and isotope analyzer combines high-precision and low drift O2 concentration measurements with analysis of δ 18O in ambient air, making it an ideal choice for challenging applications including atmospheric oxygen monitoring, to determine biogeochemical processes involved in the carbon cycle.
Product Features
Measure the concentration of O2 gas and δ 18O in the atmosphere
Two measurement methods: O2 concentration only mode and δ 18O+O2 concentration mode
Accuracy less than 2 ppm in O2 concentration mode
The maximum peak to peak drift at standard temperature and pressure is 6 ppm
Equipped with water measurement and calibration functions
Main technical parameters
|
Picarro G2207-iPerformance specifications |
|
||
|
[O2]pattern |
|
||
|
Dry under environmental concentration[O2]accuracy (1-σ,5 second/ 5minute,21% O2 Below) |
< 20 ppm / < 2 ppm |
||
|
Under standard temperature and pressureO2 Maximum drift at time(24 Within an hour, 1 The difference between the maximum values of hourly averages, in21% O2 Below) |
< 6 ppm |
||
|
[O2]Scope of work |
5–25% |
||
|
[H2O]Accuracy(1-σ,5 Seconds) |
5 ppm + 0.1%reading |
||
|
[O2] + δ18Opattern |
|
||
|
At ambient temperatureδ18Oaccuracy (1-σ,5 second/ 5Minutes) |
< 8‰ / < 1‰ |
||
|
Under environmental concentration[O2]accuracy (1-σ,5 second/ 5Minutes) |
< 300 ppm / < 30 ppm |
||
|
Under standard temperature and pressureδ18OMaximum drift(24 Within an hour,1 The difference between the maximum values of hourly averages |
< 2‰ |
||
Picarro G2207-iAnalyzer specifications |
|
|
||
measuring technique |
Cavity oscillation spectroscopy( CRDS )technology |
|
||
Measurement pool temperature control |
±0.005 ℃ |
|
||
Measurement pool pressure control |
±0.0002Atmospheric pressure |
|
||
Shock and vibration testing |
Satisfy military requirements for impact and vibrationMIL-STD 810FTesting standards |
|
||
Sample flow rate |
stay760 Supported by80–110Standard milliliters per minute( sccm ) |
|
||
Sample temperature |
-10 to45 ℃ |
|
||
Sample pressure |
300 to 1000 Tuo(40 to 133 Kilopascals) |
|
||
Sample humidity |
relative humidity( RH )less than99%(In)40 ℃Under non condensing conditions, there is no need for drying |
|
||
ambient temperature |
10 to35 ℃(Running)-10 to50 ℃(Storage) |
|
||
ambient humidity |
relative humidity( RH )less than99%Under non condensing conditions |
|
||
enclosure |
Pump (external), keyboard, mouseLCDMonitor (optional) |
|
||
data output |
RS-232EthernetUSBSimulation (optional)0–10 V |
|
||
Air inlet connector |
¼inchSwagelok ® |
|
||
External dimensions |
Analyzer:17 Inches widex 7Inch heightx 17.6Inch length(43.2 x 17.8 x 44.6Centimeters), excluding 0.5 Inch foot external pump:5.6 Inches widex 6.4Inch heightx 11.9Inch length(14.3 x 16.3 x 30.3Centimeters) |
|
||
Installation form |
Workbench or 19 Inch rack mounted chassis |
|
||
weight |
27.4 Kilogram(60.4 Pound, including external pump |
|
||
Power requirements |
100–240Alternating current,47–63 Hz(Automatic detection), when started (total) less than 260 Tile; 125 Watt (analyzer), steady state is 80 Watt (pump) |
|