Picarro G2301 high-precision gas concentration analyzer
Picarro G2301 high-precision gas concentration analyzer
Product details

Product Overview

The Picarro G2301 gas concentration analyzer can simultaneously measure carbon dioxide (CO2), methane (CH4), and water vapor (H2O), with a sensitivity of parts per billion (ppb). The drift generated for atmospheric science, air quality, and quantitative emission applications can be ignored. This analyzer meets the performance requirements of the World Meteorological Organization (WMO) and the Integrated Carbon Observing System (ICOS) for atmospheric monitoring of CO2 and CH4.

Product Features

Simultaneous and continuous measurement of three atmospheric trace gases

Billion parts per billion (ppb) sensitivity, accuracy, and precision, with extremely low drift

Complies with the international environmental and atmospheric monitoring requirements of WMO and ICOS

Equipped with water correction function, automatically reporting dry gas mole fraction

Main technical parameters

Picarro G2301Environmental, impact, and vibration tests (conducted on each analyzer)

Thermal temperature change test (whole instrument)

exceed 12 Test the working temperature step by step for hours, starting from5 ℃reach40 ℃Every increase5 ℃Maintain afterwards 40 minute.

Verify performance specifications throughout the entire testing scope.

Long term thermal testing (entire instrument)

The working temperature is tested step by step for a week, starting from30 ℃reach40 ℃Every increase5 ℃Maintain afterwards 40 minute.

Verify performance specifications throughout the entire testing scope.

Storage testing (whole instrument)

stay-10 ℃and50 ℃Constant temperature storage (not working)+Performance confirmation after testing

Vibration testing

Two axes,25 Hz1gp-pAcceleration, per axis 15 minute

useMIL-STD 810FConduct impact testing

Lift the analyzer to one side4Inch height, then release it to drop onto a hard surface (test bench).

All of them 12 Edge(xyzPerform the impact test on all axes, with each edge tested twice

Power interruption test

at least 5 Successful AC power outage and restart

Picarro G2301Ensure performance specifications (in air)

CO2

CH4

H2O

Accuracy(1-σ: Initial 5 second/ 5Ensure accuracy of minute average data under the following specified operating conditions

< 70 ppb / 25 ppb

< 0.5 ppb / 0.22 ppb

< 80 ppm / 30 ppm

Maximum drift under standard temperature and pressure(24 hour/Monthly)

*50 The difference between the maximum values of the minute averages

Ensure drift under the following specified operating conditions

< 120 ppb / 500 ppb

< 1 ppb / 3 ppb

< 100 ppm ±0.5%reading

Automatic determination of dry mole fraction

Already included

Already included

Not applicable

measuring range

01000 ppm

020 ppm

07 %v39 ℃Dew point)

Under non condensing conditions

Ensure accuracy range

300700 ppm

13 ppm

03 %v25 ℃Dew point)

Under non condensing conditions

Measurement interval (data acquisition rate)

< 5second

< 5second

< 5second

Gas response: rising/Descent time( 1090% / 9010% )

< 3second

< 3second

< 3second

Measurement pool control

Temperature:±0.005 ℃Pressure:±0.0002Atmospheric pressure