The Picarro G2301 gas concentration analyzer can simultaneously measure carbon dioxide (CO2), methane (CH4), and water vapor (H2O), with a sensitivity of parts per billion (ppb). The drift generated for atmospheric science, air quality, and quantitative emission applications can be ignored. This analyzer meets the performance requirements of the World Meteorological Organization (WMO) and the Integrated Carbon Observing System (ICOS) for atmospheric monitoring of CO2 and CH4.
Product Features
• Simultaneous and continuous measurement of three atmospheric trace gases
• Billion parts per billion (ppb) sensitivity, accuracy, and precision, with extremely low drift
• Complies with the international environmental and atmospheric monitoring requirements of WMO and ICOS
• Equipped with water correction function, automatically reporting dry gas mole fraction
Main technical parameters
Picarro G2301Environmental, impact, and vibration tests (conducted on each analyzer) |
|
|||
Thermal temperature change test (whole instrument) |
exceed 12 Test the working temperature step by step for hours, starting from5 ℃reach40 ℃Every increase5 ℃Maintain afterwards 40 minute. Verify performance specifications throughout the entire testing scope. |
|
||
Long term thermal testing (entire instrument) |
The working temperature is tested step by step for a week, starting from30 ℃reach40 ℃Every increase5 ℃Maintain afterwards 40 minute. Verify performance specifications throughout the entire testing scope. |
|
||
Storage testing (whole instrument) |
stay-10 ℃and50 ℃Constant temperature storage (not working)+Performance confirmation after testing |
|
||
Vibration testing |
Two axes,25 Hz,1gp-pAcceleration, per axis 15 minute |
|
||
useMIL-STD 810FConduct impact testing |
Lift the analyzer to one side4Inch height, then release it to drop onto a hard surface (test bench). All of them 12 Edge(x、y、zPerform the impact test on all axes, with each edge tested twice |
|
||
Power interruption test |
at least 5 Successful AC power outage and restart |
|
||
Picarro G2301Ensure performance specifications (in air) |
CO2 |
CH4 |
H2O |
|
Accuracy(1-σ: Initial 5 second/ 5Ensure accuracy of minute average data under the following specified operating conditions |
< 70 ppb / 25 ppb |
< 0.5 ppb / 0.22 ppb |
< 80 ppm / 30 ppm |
|
Maximum drift under standard temperature and pressure(24 hour/Monthly) *(50 The difference between the maximum values of the minute averages Ensure drift under the following specified operating conditions |
< 120 ppb / 500 ppb |
< 1 ppb / 3 ppb |
< 100 ppm ±0.5%reading |
|
Automatic determination of dry mole fraction |
Already included |
Already included |
Not applicable |
|
measuring range |
0–1000 ppm |
0–20 ppm |
0–7 %v(39 ℃Dew point) Under non condensing conditions |
|
Ensure accuracy range |
300–700 ppm |
1–3 ppm |
0–3 %v(25 ℃Dew point) Under non condensing conditions |
|
Measurement interval (data acquisition rate) |
< 5second |
< 5second |
< 5second |
|
Gas response: rising/Descent time( 10–90% / 90–10% ) |
< 3second |
< 3second |
< 3second |
|
Measurement pool control |
Temperature:±0.005 ℃Pressure:±0.0002Atmospheric pressure |
|