Product Overview
The Picarro G5131-i isotope and gas concentration analyzer can simultaneously measure site specificity and batch δ 15N and δ 18O in N2O. It is an ideal solution for identifying and measuring N2O emission sources in real-time on-site or in the laboratory by collecting samples. The role of N2O isotopes is to detect nitrogen sources and sinks in the global nitrogen cycle by identifying nitrification and denitrification processes in soil and water.
Product Features
• Realize high-precision measurement under atmospheric concentration
• Measurement of δ 15N compound specificity and site specificity
• Measurement of δ 18O
• No refrigerant, continuous operation, can be deployed on-site and in the laboratory
Main technical parameters
Picarro G5131-iPerformance specifications |
grid |
|
|
|
|
Target components |
accuracy1-σ 10 Minute average |
accuracy1-σ 300 Second average value |
concentration range (In the air)N2OThe unit isppb) |
Maximum drift 24 Within an hour,1 The difference between the maximum values of hourly averages |
|
N2O(Concentration) |
< 0.05 ppb |
< 0.1 ppb |
300–1500 |
< 0.2 ppb |
|
δ15N、δ15Nα、δ15Nβ |
< 0.7‰ |
< 1‰ |
300–1500 |
< 3‰ |
|
δ18O |
< 0.7‰ |
< 1‰ |
300–1500 |
< 3‰ |
|
Picarro G5131-iSystem specifications |
|
||||
measuring technique |
Cavity oscillation spectroscopy(CRDS)Technology |
||||
Measurement interval |
< 10second |
||||
response time (10%–90%) |
30 Standard milliliters per minute(sccm)Less than 30 second |
||||
Temperature sensitivity 330 ppmFunction of ambient temperature |
N2OConcentration:< 0.005 ppb / ℃The typical value is0.001 ppb / ℃) N2OIsotope:< 0.1‰ / ℃ |
||||
data output |
RS-232EthernetUSB |
||||
Air inlet connector |
¼inchSwagelok ® |
||||
External dimensions (dual box system) |
17 Inches widex 12Inch heightx 27Inch length (43 x 32 x 69Centimeters) |
||||
weight |
87 Pound(40 Kilogram) |
||||
Power requirements |
When turned on, it is 300 Watt, in steady state is 210 tile |
||||
Picarro G5131-iOperating conditions |
|
||||
Sample temperature |
-10 to45 ℃ |
||||
Sample flow rate |
stay 760 Underneath less than 50 Standard milliliters per minute(sccm)No need for filtering |
||||
Sample pressure |
300 to 1000 Tuo(40 to 133 Kilopascals) |
||||
Sample humidity |
0–2% v H2O(18 ℃Dew point), under non condensing conditions |
||||
ambient temperature |
15 to35 ℃(Running) -10 to50 ℃(Storage) |
||||
ambient humidity |
Relative humidity(RH)Less than99%Under non condensing conditions |
||||
System transportation |
Not herePicarroThe warranty right will be cancelled for transportation in the shipping box |
||||
interfere |
This instrument is designed to measure specific gases in ambient air or air like matrices. such asCOandCH4 Other high concentration gases may interfere with the measurement of the instrument. For more details and suggestions, please contactPicarro. |