Product Overview
The Picarro G5310 gas concentration analyzer can synchronously and accurately measure nitrous oxide (N2O), carbon monoxide (CO), and water vapor (H2O), with a sensitivity of trillions of parts per trillion (ppt). The drift generated by atmospheric science, air quality, and quantitative emission applications can be ignored. This analyzer meets the performance requirements of the World Meteorological Organization (WMO) and the Integrated Carbon Observing System (ICOS) for atmospheric monitoring of N2O and CO.
For N2O and CO, the accuracy at 5 seconds and 5 minutes is<200 ppt and<400 ppt, respectively. For N2O and CO, the drift within 24 hours at standard temperature and pressure (STP) is less than 100 ppt. The water calibration software will automatically report the dry gas mole fraction to help reduce the complexity of research work and save consumable costs.
Product Features
• Synchronous continuous measurement of N2O and CO
• Mid infrared cavity ring down spectroscopy (CRDS) for high-precision and low drift analysis
• Complies with the international environmental and atmospheric monitoring requirements of WMO and ICOS
• Equipped with water correction function, automatically reporting dry gas mole fraction
Main technical parameters
G5310performance index |
N2O |
CO |
H2O |
|
Accuracy(1σ,5 Accuracy in seconds(1σ,5 Minute accuracy(1σ,1 Hour) |
< 0.2 (N2O < 500 ppb) < 0.04 ppb (N2O < 500 ppb) < 0.02 ppb (N2O < 500 ppb) |
< 0.2 (CO < 400 ppb) < 0.04 ppb (CO < 400 ppb) < 0.02 ppb (CO < 400 ppb) |
40 ppm 6 ppm - |
|
Accuracy(1σ,5 Minutes) |
0.03 %ofN2Oreading(N2O > 500 ppb) |
0.02 %ofCOreading(CO > 400 ppb) |
6 ppm |
|
Drift(24 Hour) |
< 0.1 ppb |
< 0.1 ppb |
- |
|
measuring range |
1 – 1500 ppb |
1 – 1500 ppb |
0 – 3% |
|
Measurement interval |
< 5second |
< 5second |
< 5second |
|
G5310System indicators |
|
|||
measuring technique |
Cavity ring down spectroscopy(CRDS) |
|||
Measurement pool temperature control |
±0.005℃ |
|||
Measurement pool pressure control |
±0.0002Atmospheric pressure |
|||
Sample temperature |
-10 to +45 ℃ |
|||
Sample pressure |
300 to1,000Tuo(40 to 133 Kilopascals) |
|||
Sample flow rate |
100 Standard cubic centimeters per minute(sccm) |
|||
Sample humidity |
< 99%Relative humidity (in)40 ℃Under non condensing conditions, there is no need for drying |
|||
Environmental temperature range |
15 to 35 ℃(Working conditions) -10 to 50 ℃(Storage conditions) |
|||
ambient humidity |
< 99%relative humidity(Under non condensing conditions) |
|||
enclosure |
Keyboard, mouse, LCD display (optional) |
|||
data output |
RS-232, in order toTai Wang,USB |
|||
Pipe joint |
¼inchSwagelok ® |
|||
size |
17Inches widex 12Inch heightx 27Inch deep(43 x 32 x 69centimeter) |
|||
Installation form |
Workbench |
|||
weight |
95 Pound(43 Kilograms), Including built-in pump |
|||
power waste |
100 – 240Alternating current,47 – 63Hz(Automatic)Induction)When powered on 300 WAt steady state210 W |