Instrument Introduction:
The practical total phosphorus rapid analyzer is developed and designed according to the standard HJ/T399-2007. Peroxysulfate is added to the water sample to convert all phosphorus in the solution into orthophosphate. The orthophosphate in the sample reacts with ammonium molybdate in an acidic medium in the presence of antimony salt to form phosphomolybdic acid, which is immediately reduced by ascorbic acid to form a blue complex. After measuring its absorbance, the instrument directly displays the total phosphorus content in mg/L.
Technical parameters:
Measuring Item |
total phosphorus |
Measurement range |
0-20mg/L |
accuracy |
±5% |
detection limit |
0.02mg/L |
Optical stability |
≤0.001A/20minute |
Measurement time |
30minute16Branch sample/write instructions or comments on |
Temperature indication error |
±2℃ |
Temperature uniformity |
±1℃ |
Light source lifespan |
1010000 hours |
Number of curves |
standard3strip extend77strip |
storing data |
1800individual |
display |
LCDdisplay |
Display interface |
All in Chinese |
Colorimetric method |
Colorimetric tube |
printer |
Thermal printer |
data interface |
USB |
ambient temperature |
5-40℃ |
ambient humidity |
Relative humidity ≤85% |
safety protection |
Dual security protection |
Pre set digestion mode |
CODTotal phosphorus and total nitrogen |
Reminder alarm |
buzzer |
Customer customization |
support |
rated voltage |
AC220V±10%/50Hz |
rated power |
instrument<10W, digestion apparatus<350W |
Alarm prompt1 |
Reaching the specified temperature and maintaining a constant temperature |
Alarm prompt2 |
Dissolve completed and stop heating |
Instrument dimensions |
310mm*230*150 |
Dimensions of the digestion instrument |
230*340*130(mm) |
Product features:
1. Relevant national standardsHJ/T399-2007 R&D DesignAccurate, fast, and effective data detection.
2. Imported cold light source is used, with excellent optical performance, good stability, and a light source lifespan of 100000 hours.
3. 5-inch LCD screen, full Chinese display, direct data reading, simple and time-saving operation.
4. The practical integrated optical colorimetric tube for digestion detection does not require pipetting or tube replacement, making the measurement simple, fast, and safe.
5. Built in 80 curves and 1800 measurement data, which can be expanded by customers themselves.
6. The system comes with a one click recovery setting, which can quickly recover in case of data loss due to misoperation.
8. It has data storage function and power-off protection function, making it convenient to query historical data and prevent data loss.
9. External USB interface, data can be transferred to the computer for long-term storage.
10. Embedded printer that can immediately print detection data or save records for printing.
11. The digestion instrument adopts intelligent PID temperature control technology and a dual anti overheating protection system, with safe and uniform heating and fast speed, and is suitable for the digestion of COD, total phosphorus, total nitrogen and other projects.
Standard configuration:
Serial Number |
name call |
quantity |
Serial Number |
name call |
quantity |
1 |
Practical total phosphorus rapid analyzer host |
1tower |
6 |
Simplified operation flowchart |
1share |
2 |
Colorimetric tube |
10branch |
7 |
operation instructions |
1share |
3 |
Color tube cleaning cloth |
1piece |
8 |
certificate of conformity\\warranty card |
1share |
4 |
Total phosphorus detection reagent |
1cover |
9 |
Power Cables |
1strip |
5 |
USBdata line |
1strip |
10 |
Drive CD |
1fix |