characteristic
Excellent long-term stability;
The temperature curve above 100K shows a linear response;
Better resistance to radiation effects.
parameter
※ Temperature range: 1.2K~325K;
※ Standard curve: None
Recommended excitation current: 1mA
※ Recommended dissipated power under incentive:3μW@4.2K, 50μW@273K
※ Response time: 10S@273K
Radiation impact: Can be used in ionizing radiation environments
※ Magnetic field influence: Not recommended for use in magnetic field environments below 77K temperature
Repeatability: ± 5mK@77K (Based on 100 thermal shock tests from 300K to 77K)
Typical curve characteristics
Calibration accuracy and range
model
calibration accuracy
calibration range
RI20-BC-1.2L
0.1K
1.2K-300K
RI20-FK-1.2L
0.1K
1.2K-300K
Accuracy of thermometer=(Calibration uncertainty) 2 +repeatability 2)0.5
Typical temperature and resistance relationship
T(k)
R(Ω)
dR/dT(Ω/K)
1.2K
2.3674
0.2802
4.2K
3.1602
0.2319
20K
5.251
0.0771
77K
11.898
0.1802
150K
26.308
0.1986
273K
50
0.1907
External dimensions and packaging dimensions
Size:Φ2.2mm*20mm
Quality:500mg
Lead wire:2root
Bare chip packaging form: rhodium iron alloy encapsulated inside the ceramic, Externally encapsulated within an alumina shell
physical property