The 2600 series spectrophotometer has advanced functions such as quantitative measurement, wavelength scanning, time scanning, multi wavelength testing, DNA protein testing, etc. It has the characteristics of high precision, high resolution, low noise, and low stray light, and is particularly suitable for fields such as scientific research, pharmaceutical inspection, life sciences, environmental monitoring, metallurgy and chemical engineering.
Instrument functions
◆ Luminescence measurement: transmittance T%, absorbance Abs, energy E
Quantitative testing: concentration content C (coefficient method, standard curve method)
◆ Wavelength scanning (spectral scanning): scanning range, interval, and speed can be set
Time scanning (kinetics): measure the change spectrum and calculate the reaction rate
◆ DNA/protein determination: ratio and concentration are directly displayed, and customizable methods are available
◆ Multi wavelength testing: simultaneously measuring results at multiple arbitrary wavelengths
Other functions: system application, graph data processing, report printing.
Instrument characteristics
◆ Dual beam optical path, low drift and high stability, capable of compensating for reference changes, suitable for long-term testing
◆ Push pull sample cover design, can be opened and closed with both hands, easy to operate
◆ Spacious sample room, capable of accommodating 5-100mm colorimetric dishes, supporting various accessory parts required for testing
High quality optical and electronic components ensure high precision and low stray light of the instrument
◆ Large screen dot matrix LCD, clear display of graph data, user-friendly interface, easy operation
◆ Smaller spectral bandwidth, higher resolution, fully meets pharmacopoeia requirements, and has a wider range of applications
◆ Long life environmentally friendly flange seat deuterium lamp, no optical debugging required for lamp replacement
Real time saving of graphs and data, with the ability to maintain power off, printable, and exportable to PC
◆ Sturdy base plate and optical bracket, imported key components, durable and long-lasting
Standard PC scanning software, more convenient online operation, and expanded data processing and storage capabilities
Technical indicators and basic parameters
model |
S-2600CRT |
spectral bandwidth |
0.5/1/1.5/2/4/5nmadjustable |
wavelength range |
190~1100nm |
optical system |
Dual beam optical path, holographic diffraction grating |
wavelength accuracy |
±0.1nm(656.1nm)± 0.3nm (full area) |
Wavelength Repeatability |
≤0.1nm |
Wavelength setting |
Automatic wavelength setting, wavelength resolution of 0.1nm |
Photometric range |
0~200%T, - 4~4A, 0~9999C, 0~9999F |
photometric accuracy |
±0.002A (0~0.5A), ±0.004A (0.5~1A), ±0.2%T (0~100%T) |
Photometric Repeatability |
≤0.001A (0~0.5A),≤0.002A (0.5~1A), ≤0.1%T (0~100%T) |
stray |
≤0.03%T (220/360nm) |
Baseline drift(Stability) |
≤0.0004A/h (500nm,0A) |
noise |
±0.1%T(0%TLine); ± 0.3% T (100% T-line); ± 0.0004A (at 500nm) |
baseline flatness |
±0.001A |
scanning speed |
Three adjustable speeds: high, medium, and low |
detector |
Imported silicon photodiode |
light source |
Imported long-life tungsten and deuterium lamps(No optical debugging required |
display |
6320x240 inchesLCD |
keyboard |
Digital touch film button |
data interface |
USBopening |
print out |
Micro printer; PCPrinter (for online use) |
PCsoftware |
PCScanning analysis software(Standard configuration) |
power supply |
AC90-250V, 50/60Hz |
External dimensions |
635x515x255mm |
weight |
28kg |