Introduction to SDW-160 Portable Oil Particle Counter
Adopting the shading counting principle developed by the International Hydraulic Standards Committee
Built in multiple commonly used standards such as GJB 420, NAS1638, ISO4406, SAE4059E, etc
Adopting advanced metering pump injection system, supporting custom volume
Main applications
Laboratory oil analysis
On site and online system oil analysis
Acceptance of hydraulic equipment and daily maintenance
■ System cleaning verification
■ Cleaning verification of components
Wear test of hydraulic components
SDW-160 portable oil particle counterIt is a specialized instrument developed based on national and international standards such as GB/T 18854-2002 (ISO11171-1999) for detecting pollution levels in oil. Suitable for on-site and laboratory pollution testing of hydraulic oil, lubricating oil, shale oil, transformer oil (insulation oil), turbine oil (turbine oil), gear oil, engine oil, aviation kerosene, water-based hydraulic oil, phosphate ester oil and other oils.It can also be installed on various hydraulic transmission, oil filter, cleaning machine, testing bench and other systems to achieve online detection of system oil cleanliness.It can be widely applied in fields such as aerospace, power, petroleum, chemical, transportation, ports, metallurgy, machinery, automobile manufacturing, etc.
main features
The principle of light resistance (shading) test specified by the International Hydraulic Standards Committee
High precision laser sensor with a wide testing range, stable performance, low noise, and high resolution
High precision metering pump sampling method, adjustable injection speed, high sampling volume accuracy
The pipeline is made of 316L and PTFE materials to meet the testing requirements of various organic solvents and oil products
■ Used for laboratory or on-site measurement, with built-in pressure reducing device for online high-pressure measurement
■ It can be connected to an external pressure chamber to form positive/negative pressure, enabling the detection and degassing of high viscosity samples
Various sampling containers such as standard sampling bottles and cups can be used, or directly connected to hydraulic systems to meet the testing requirements of different industries
Built in multiple calibration curves, compatible with all commonly used domestic and international standards for calibration
■built-inGJB-420A、GJB-420B、NAS1638、ISO4406、SAE4059EAnd ГOCT17216Waiting for multiple commonly used standards, supporting custom standard testing, and setting required standards according to customer needs
Can set up 1000 particle size channels for particle size analysis
■ Built in data analysis system that can automatically determine sample grade based on standards, with automatic data processing and printing functions
■ Color touch screen operation, Chinese input, with preset, input, modification, and storage functions, easy and fast operation
Equipped with RS232 interface, it can be connected to a computer or laboratory platform for data processing, and UBS can also be used for data storage
■ Built in lithium battery, can be used without external power supply
TECHNICAL INDEX
■ Light source: semiconductor laser
■ Particle size range: 0.8-500um
■ Detection channels: 8 channels, freely set particle size
Sampling volume: 0.2-1000ml
Sampling accuracy: better than ± 1%
Sampling speed: 5-80mL/min
Counting accuracy: ± 10%
Resolution: ≤ 10%
■ Repeatability: RSD<2%
■ Limit coincidence error: 10000 particles/mL
Pressure range: High pressure 5-420bar, low pressure 0-6bar
■ Viscosity range: ≤ 350cSt
■ Power supply: AC100-240V,50/60Hz
■ Battery capacity: 5200mAh
■ Running time: 6-8 hours
■ Dimensions: 310 × 305 × 135mm