SYD-3536AThe automatic open flash point tester is based on the standards of the People's Republic of ChinaGB/T 3536-2008Designed and manufactured in accordance with the requirements specified in the "Cleveland Open Cup Method for Determination of Flash Point and Ignition Point of Petroleum Products", it is applicable to the determination of fuel oil and open flash point below79The Klimov open cup method flash point of petroleum products at ℃.
SYD-3536AThe Type Automatic Open Flash Point Tester is an automatic open flash point tester that is an ideal substitute for similar imported instruments. It can be widely used in the railway, aviation, power, petroleum industries, and scientific research departments.
1、 Main technical features
1Simulate and track the function curve of temperature rise and test time, with Chinese misoperation software prompt modification function, equipped with test
Parameter prompt function for testing date, testing time, etc.
2Automatically calibrate the influence of atmospheric pressure on the experiment and calculate the correction value.
3Differential detection, automatic correction of system deviations.
4The scanning, detection, and printing of data are automatically completed, and the test arm is automatically raised and lowered.
5Automatic control of heating rate, forced air cooling of the heating furnace.
6Humanized appearance design, beautiful, safe, and easy to operate.
2、 Main technical parameters and indicators
1Working power supply:AC220(-10%~+5%)V、50Hz;
2Temperature measurement: Range:50~400℃;
Repeatability:≤8℃;
Reproducibility:≤16℃;
Temperature display unit (resolution):0.1℃;
Accuracy:0.5%
3Heating rate: in accordance withGB/T3536Standard;
4Ignition method: Electronic ignition gun ignition
(Gas flame)3~4mmIgnite);
5Environmental temperature:(10~40)℃;
6Relative humidity: ≤80%;
7Power consumption of the whole machine: not greater than500W.