Shenzhen Environmental Radon DetectorApplication field
Air Radon Measurement (Standard)
Soil Radon Measurement (Standard)
Radon measurement in water (optional)
Radon emission measurement (optional)
Shenzhen Environmental Radon DetectorMain technical performance
1.Sensitivity: ≥ 0.68cpm/[Bq • m-3]
2.Background count rate: ≤ 0.3cpm
3.measurement range
Environmental air radon: (3-100000) Bq/m ³
Soil radon: (300-300000) Bq/m ³
Radon release rate: (0.001~10.000) Bq/[m2•s]
Radon in water: (0.003-100) Bq/L
4.Measurement repeatability error: ≤ 5% (radon chamber concentration 2000Bq/m ³, environmental humidity 65%, temperature 25 ℃).
5.Long term stability (8h) error: ≤ 10%
6.Power supply: Lithium ion rechargeable battery/AC power, battery powered can work continuously for 30 hours.
7.working environment
Temperature: -10 ℃ to+40 ℃
Humidity: Relative humidity ≤ 90% (+40 ℃)
8.Detector: Zinc sulfide ZnS (Ag) and photomultiplier tube combination system
9.Data storage: capable of storing 2000 pieces of data
10.Operation mode: single point detection or continuous monitoring
11.Display: LCD liquid crystal display
12.Gas extraction method: active pump suction
13.measure time
Air radon: 31 minutes
Soil radon: 11 minutes
Radon release rate: 30 minutes
Radon in water: 31 minutes
14.Print data: date, time, point number, and test results
15.Size: (330 × 210 × 170) mm
16.Weight: 5kg (main unit)
The instrument complies with the measurement principles and requirements of GB50325-2001 "Code for Indoor Environmental Pollution Control of Civil Building Engineering" (2006 edition) and GB/T16147-1995 "Measurement Method for Radon Concentration in Air by Scintillation Bottle".
Instrument Features
1.Small in size, lightweight, and easy to carry.
2.High sensitivity, low power consumption, dual-use of AC and DC, DC power supply can support instrument operation for 30 hours.
3.USBData transmission interface, Bluetooth printing.