Name:Silent dry air generator
Model:GA-2009
Manufacturer:Guangrun Instrument Equipment Company
Product Introduction:
The air used in the laboratory has strict requirements to ensureGas chromatography analysisWhen the experiment proceeds normally, it is necessary to process the air we come into contact with in daily life in order to be efficiently used, thus "giving birth" to a dedicated systemAir generatorThe air generator has replaced the traditional high-pressure air cylinder.Silent, dry, oil-free air generatorIt is generally composed of an air pump/compressor, a pressure stabilization system, a pressure control system, a purification system, and a display system.
Technical parameters:
1. Output pressure: 0.4-0.5MPa
2. Output flow rate: 2000/5000 (ml/min)
3. Air purity: oil-free grade three
4. Actual dimensions: 230 * 480 * 350 (mm)
Main features:
1. Three level purification ensures clean air, with a dual pressure stabilization valve and buffer tank for stable output pressure

Model:GA-2009
Manufacturer:Guangrun Instrument Equipment Company
Product Introduction:
The air used in the laboratory has strict requirements to ensureGas chromatography analysisWhen the experiment proceeds normally, it is necessary to process the air we come into contact with in daily life in order to be efficiently used, thus "giving birth" to a dedicated systemAir generatorThe air generator has replaced the traditional high-pressure air cylinder.Silent, dry, oil-free air generatorIt is generally composed of an air pump/compressor, a pressure stabilization system, a pressure control system, a purification system, and a display system.
GA-2009 Silent Oil Free Air GeneratorPowered by a fully enclosed compressor, natural air is purified through three stages to remove moisture, oil, and impurities from the air. Stable and clean air is then output through a stabilizing device.GA-2009 Silent Oil Free Air GeneratorIt has the characteristics of stable output pressure, low noise, and strong purification ability, making it an ideal air source to replace high-pressure air cylinders. This product can not only meet the needs of various domestic and imported modelsgas chromatographAnd the use of various analytical laboratories, which can also be used asHigh purity nitrogen generatorThe air source.
GA-2009 Silent Oil Free Air GeneratorNo noise and easy to install, can continuously generate zero level air, and uses modern filtration technology to remove particles, water, and oil mist. Completely remove hydrocarbons such as catalytic cracking methane. The maximum flow rate and temperature of the carrier gas/gas can be selectively controlled, and the power supply can also be monitored.Oil-free air generatorInterruption or timeout alarm, in compliance with existing international standards. Upon detecting a malfunction, the automatic switching device option can be selected to automatically switch to the backup gas cylinder or cut off the gas delivery to the laboratory. Subsequently, the user can eliminate the malfunction and reset the alarm. Eliminating expensive and dangerous gas cylinders for laboratory supply.
GA-2009 Silent Oil Free Air GeneratorNo noise and easy to install, can continuously generate zero level air, and uses modern filtration technology to remove particles, water, and oil mist. Completely remove hydrocarbons such as catalytic cracking methane. The maximum flow rate and temperature of the carrier gas/gas can be selectively controlled, and the power supply can also be monitored.Oil-free air generatorInterruption or timeout alarm, in compliance with existing international standards. Upon detecting a malfunction, the automatic switching device option can be selected to automatically switch to the backup gas cylinder or cut off the gas delivery to the laboratory. Subsequently, the user can eliminate the malfunction and reset the alarm. Eliminating expensive and dangerous gas cylinders for laboratory supply.
Technical parameters:
1. Output pressure: 0.4-0.5MPa
2. Output flow rate: 2000/5000 (ml/min)
3. Air purity: oil-free grade three
4. Actual dimensions: 230 * 480 * 350 (mm)
Main features:
1. Three level purification ensures clean air, with a dual pressure stabilization valve and buffer tank for stable output pressure
2. High strength purification tube, absolutely safe to use
3. Low noise, stable baseline
4. No need to purchase and store gas cylinders
5. Low voltage operation, can be freely placed in experiments
6. Greatly saves time and space
7. If the system detects a fault, it can automatically switch to the backup gas cylinder or cut off gas transmission


Simple faults and troubleshooting methods:
1. After turning on, the green working indicator light is on, and the compressor starts but the pressure does not increase.
2. If it is the first time starting after opening the box, please check the air inlet and replace the plug with a filter.
3. Block the output port, check if there is any air leakage at the joints of the drying tube and gas circuit, and seal the leaking area.
4. After turning on, the green working indicator light will turn on and the compressor will not start. Please open the right cover of the instrument and check if the thermal protection relay behind the compressor is burnt out. If so, please replace it.
5. After the air generator is turned on, the green working indicator light does not light up, the compressor does not start, and there is no output pressure. Please open the right cover of the instrument and check whether the transformer is working properly.