Rapid investment return
• Optimize deodorization process
• Ensure compliance with the contract
• Automatic sealed oil delivery verification
• Avoid contamination of pipeline "locking ports"
Measurement certainty
Multiple streams with up to 8 channels
A single calibration gas for all sulfur species
• Bipolar response at ppb to% levels
• Performance is not affected by changes in hydrocarbons, process equipment, or mixing
Provide services based on customer needs. Differentiated sulfur measurement for precise process analysis. Do you need to monitor the concentration of odorants in hydrocarbon gas? This application requires highly sensitive detection and measurement of individual sulfur components. The application of transfer pipes in some natural gas pipelines also requires detailed reports on the concentration of sulfur components. GALVANIC's SULFURCHROME online chromatography instrument enables easy and real-time measurement of all chemical components containing sulfur up to a concentration of 5ppb.
SulfurChrome ™ GC product specifications
Performance of Total Sulfur Analyzer:
|
|
accuracy
|
± 2% full range
|
Repeatability
|
± 2% full range
|
sensitivity
|
± 2% full range
|
response time
|
T90: 4~15 minutes, with an average of 6 minutes [specific application] |
Method/Technology
|
Gas chromatography/sulfur chemiluminescence detector
|
Sulfur measurement
|
H2 S, COS, CS2 , Sulfides, Disulfides, Disulfides, Thiols, Thiophenes, Total Sulfur, Others
|
Range
|
0 ~ 10 mg/l, 0~5 g/l [equimolar response to all sulfur species]
|
environment
|
|
ambient temperature
|
0 ~ 50°C [32° to 122°F]
|
relative humidity
|
0~95% non condensing
|
install
|
indoor
|
size
|
711 x 1,016 x 406 [28” x 40” x 16”]
|
weight
|
57 kg. [125 lbs.]
|
Shell
|
Painted steel
|
Basic conditions:
|
|
Power/Consumption
|
100~240 VAC, 600 watts
|
helium
|
10 cc/min, 4 bar [60 psi]
|
hydrogen
|
80 cc/min, 1 bar [15 psi]
|
Calibration gas
|
H2S in nitrogen [for all sulfur species]
|
Vacuum pump
|
115 VAC, 6 Amp consumption
|
Communication:
|
|
Analog output
|
4 x 4-20mA output (cyclic or automatic)
|
Analog input
|
3 x 4-20mA input [RTD, 4-20mA, transducer] |
Modbus
|
TCP/IP, RS232, Or RS485 |
Digital input
|
4 Discrete inputs
|
Digital output
|
4 x SPDT relays
|
interface
|
Color LCD display and expandable keyboard
|
PC GUI
|
Ethernet access
|
CEC / NEC
|
Class I Div. 2 Groups BCD T3 [Purged]
|