advantage
◎ Over range automatic dilution reanalysis.
◎ Automatic leakage detection alarm.
The Chinese version of the touch screen operation interface is simple to operate and maintain.
◎ With self diagnostic function, it can identify whether there is a lack of water samples or chemicals
Long term self-control, low maintenance, and low operating costs
Low medication consumption and short preparation time
◎ After power failure, it has a self starting function for incoming calls
Real time display of measurement progress.
Dual light path detection, real-time display of absorbance.
◎ Automatic compensation for turbidity/chromaticity interference.
◎ UV+high temperature and high pressure dual path digestion, higher efficiency
The calibration method is flexible, and the calibration points and time can be set.
Supports multiple interface designs such as USB, TF card, miniUSB, etc. for convenient data access.
Low power product design, energy-saving and environmentally friendly
◎ Automatic cleaning function
Measurement principle:
After filtering the sampleAfter being pumped into the digestion tube, potassium dichromate, silver sulfate, sulfuric acid, and mercury sulfate were added and heated to 175 ℃. After complete digestion, the analyzer measured the OD value of the generated green substance at 460nm and calculated the COD concentration in the sample based on the correction factor stored in the analyzer.
Technical Specifications:
Measurement principle: Potassium dichromate method for high-temperature digestion
Detector:460nm dual path photometer
Measurement type: User manual or set interval automatic analysis
Measurement interval: can be freely programmed according to the actual situation
Measurement method: Continuous automatic measurement
Measurement range:0~50/100/500/1000/5000mg/L,
Automatically switch range according to water quality
Detection limit:≤0.1 mg/L
Dissolving time: automatic3. 5, 10, 20, 30, 40, 60, 80, 100, or 120 minutes are optional
There are three modes to choose from:
(1) Continuous measurement mode
(2) User defined interval mode for 1-24 hours
(3) Remote start measurement mode
Repeatability:≤3%
Zero drift:±5mg/L
Range drift:±10%
Automatic cleaning, automatic full range calibration
Operating temperature:0-40 ℃, temperature: relative humidity 95%
Signal output: standard4-20mA analog output, standard RS232 digital output, capable of bidirectional communication and remote control, maximum 500 ohm load, standard Modbus output
Signal input:1-way analysis, 1-way correction
Data storage: Storage3000 sets of data, available for USB download of analysis data and operation record data
alarm1 high limit alarm, 1 regular, 1 calibration
Medication replacement:3-4 weeks, depending on the ambient temperature
Protection Level:IP55
Human computer interaction interface: Industrial grade touch screen, including Chinese version operating software Power supply:220VAC,50/60Hz
Weight:33kg (excluding medication)
Size:800x450x300mm (HxWxD)