T2009 fully automatic wet laser particle size analyzer is an intelligent fully automatic laser particle size analyzer with a testing range covering millimeter, micrometer, submicron, and nanometer full ranges. The dual beam multispectral detection system and lateral light scattering testing technology have been adopted, greatly improving the accuracy and performance of instrument testing.
Scope of application:
T2009 is widely used in cement, ceramics, pharmaceuticals, lotion, coatings, dyes, pigments, fillers, chemical products, catalysts, drilling mud, abrasives, lubricants, coal powder, silt, dust, cells, bacteria, food, additives, pesticides, explosives, graphite, photosensitive materials, fuels, ink, metal and non-metallic powders, calcium carbonate, kaolin, coal water slurry and other powdery materials.
Product features and advantages:
Fully automatic centering system | The optical path automatic alignment system is composed of a precision four item hybrid stepper motor, with micro motion accuracy reaching the micrometer level, ensuring that the instrument's optical path is always in optimal condition. |
Fully built-in decentralized system | By using a high-power circulation pump and optimizing the pipeline design, the overall control coordination has reached a very high level, effectively preventing the secondary sedimentation of large particles. |
Unconstrained free fitting technique | By using unconstrained free fitting techniques, particle size analysis is not limited by any function and can truly reflect the distribution state of particles. |
Optical path design | Adopting the converging light Fourier transform technique, the scattered light is not limited by the lens aperture; Adopting a dual spectral surface optical path design, effectively expanding the measurement range to 2000 micrometers; The original dual laser orthogonal optical path uses semiconductor assisted lasers to measure lateral scattering from 45 degrees to 135 degrees, further expanding the measurement range. |
Specification and model | T2009A | T2009B | |
Execution standards | ISO 13320-1:1999; GB/T19077.1-2008; Q/0100JWN001-2013 | ||
Range | 0.01-2000μm | 0.01-1200μm | |
Number of detector channels | 97 | 95 | |
Accuracy Error | <1% (D50 value of national standard sample) | ||
Repeatability error | <1% (D50 value of national standard sample) | ||
Laser parameters | Main laser source: High performance He Ne laser with λ=632.8nm and p>3.0mW Auxiliary laser source: blue light laser, λ= 650nm; p>2mW |
||
dispersion method | ultrasonic | Frequency: f=40KHz, Power: p=50W | |
stir | Speed: 0-3000rpm, adjustable speed | ||
cycle | Rated flow rate: 30L/min Rated power: 70W | ||
Sample Cell | Capacity: 1000mL | ||
MODE | Software operation/fully automatic operation mode can be switched | ||
Automatic centering system | Fully automatic centering system | ||
test speed | <2min/time | ||
volume | 1040mm*440mm*540mm | ||
weight | 58Kg |