TC-100 COD automatic digestion reflux analyzer
The TC-100 COD automatic digestion reflux analyzer adopts microcomputer technology for timed control of the heating electric furnace, which can simult
Product details

undefined
undefined
undefined
undefined
undefined
undefined

TC-100 COD automatic digestion reflux analyzerProduct Overview:

The instrument adopts microcomputer technology for timed control of the heating electric furnace, which can simultaneously heat 16 digestion tubes,

At the same time, it makes the use of the instrument more reassuring.

TC-100 COD automatic digestion reflux analyzerTechnical indicators:

Measurement range: 5-1000mg/L;

0-10000mg/L (dilution method);

Temperature control range: 160 ℃~185 ℃;

Dissolving time: 10-120 minutes;

Measurement error: 5-100 mg/L; Relative error ≤ ± 5 mg/L;

100 mg/L~1000 mg/L; Relative error ≤ ± 5%;

Potassium hydrogen phthalate standard solution (500mL);

Anti chlorine interference: Cl-<1000mg/L:

Dissolving time: 10-120 minutes;

Instrument features:

The digestion time and temperature can be set. After digestion is completed, the instrument automatically stops heating and can be left unattended.

The dedicated digester consists of a front panel, heater, etc. The heating furnace can be used for one-time heating and digestion16A sample.

1-1 Temperature display window: displays the current temperature value or temperature set value of the heater

1-2 Time Display Window: Display the timing time and set time value

1-3 Temperature control range: 160 ℃~185 ℃, temperature control accuracy: ± 2 ℃ (temperature control range can be adjusted from room temperature to 200 ℃ according to user requirements)

Temperature setting: The preset temperature of the heater is 165 ℃. Pressing the setting button can change the temperature setting value. After pressing the start button, the digester starts heating. eliminate

The temperature setting range is 160 ℃ to 185 ℃, and the temperature is increased by 5 ℃ every time the setting key is pressed.

1-4 Time setting: The preset time value is 15 minutes. Press the timer button to change the time setting value. After pressing the timer button, the timer function will be activated. Time setting range

The duration ranges from 10 minutes to 150 minutes, with each press of the timer button increasing the time by 5 minutes. When it reaches 150 minutes, it automatically jumps to 10 minutes.

Suggest heating the COD digester in a fume hood. Otherwise, it is recommended to cover the protective cover during heating.