UP-CYY-II Infrared Spectrometer for Oil Analysis
The UP-CYY-II infrared spectrophotometer is a dual-purpose device that uses an aluminum alloy casting base, undergoes natural failure treatment, an ex
Product details
Compliant with standards
National Standard of the People's Republic of China HJ637-2012 Water Quality - Determination of Petroleum and Animal and Vegetable Oils - Infrared photometric method

Infrared spectrophotometer oil analyzerTechnical principle:
According to the HJ637-2012 (GB/T16488-1996) standard, oil substances in water (or gas, solid) are extracted with carbon tetrachloride, and the total extract is determined. Then, the extract is adsorbed with magnesium silicate, and after removing polar substances such as animal and vegetable oils, petroleum substances are determined. The total extract and petroleum content were calculated based on the absorbance A2930, A2960, and A3030 at spectral numbers 2930 cm-1 (stretching vibration of C-H bond in CH2 group), 2960 cm-1 (stretching vibration of C-H bond in CH3 group), and 3030 cm-1 (stretching vibration of C-H bond in aromatic ring), respectively. The content of animal and vegetable oils is calculated based on the difference between the total extract and petroleum content.

Infrared spectrophotometer oil analyzerInstrument features:
One machine dual-use: using aluminum alloy casting base, natural failure treatment, external power supply, injection molded shell, beautiful and elegant, small in size and light in weight,As a laboratory instrument, it can also be used as a portable instrument in emergency monitoring.
Good stability: Adopting an integrated optical path system, the optical path design is reasonable, the signal is strong, and the signal-to-noise ratio is high. Switching power supply, wide power range.
Drift free: Using Du Chuang's automatic switching mechanism for samples and reference cells, precise positioning, error elimination, stable measurement results, and minimizing the impact of mechanical errors.
Accurate positioning: The residual cut method is used for wave number precise positioning scanning, so that the wave number positioning accuracy is less than one wave number.
True three wave number: compliant with the national standard "HJ637-2012",No need to create a standard curve, only a set of correction coefficients.
Small measurement error for different ratios: Simulate the oil composition in water and measure the error of any component standard oil, which is less than 5%, making the instrument truly serve the actual water sample.
There is a dedicated menu for measuring catering oil fume, which measures the concentration of cooking oil fume emissions in accordance with the national standard for cooking oil fume emissions (GB18483-2001).
Easy to operate: Chinese menu operation, equipped with a large LCD display, no need for sample calibration, measurement results can be printed out.It can be used independently without a computer or connected to a computer for operation, and the wave number can be freely compensated and located.
Powerful functionality: capable of both quantitative measurement and qualitative analysis; The spectrum is clear and can distinguish various interfering substances.

Infrared spectrophotometer oil analyzerTechnical parameters:
Wave number scanning range: 2400cm-1-3400 cm-1
Wave number accuracy: ± 1 cm-1
Wave number repeatability: ± 1 cm-1
Spectral resolution: 1 cm-1
Measurement range: 0.02-8000mg/L
Measurement repeatability: < 2%
Measurement accuracy: ± 2%
Linear range of absorbance: 0.0000-1.9999
Correlation coefficient: r>0.999 (infrared non dispersive method)
Zui low detection concentration: 0.002mg/L (enriched 100 times the oil concentration in water)
Detection limit 3SD:< 0.2 mg/L (solution concentration)
Baseline drift: 1%/4h
Measurement errors for different ratios:< 5%
Dimensions: 540 × 246 × 160 (mm)
Weight: 13.5kg
Power supply: 220V ± 40V 50Hz ± 1Hz 40W
Humidity:< 80%
Temperature: 5-35 ℃

Infrared spectrophotometer oil analyzerTechnical parameters:

Project TECHNICAL INDEX
Wavenumber range 2400cm-1-3400 cm-1
Wave number accuracy and wave number repeatability ±1 cm-1
Spectral resolution 1 cm-1
measuring range 0.02~8000mg/L
Measurement repeatability <2%
Measurement accuracy <±2 %
Linear range of absorbance 0.0000~1.9999
correlation coefficient r> 0.999 (Infrared Non Dispersive Method)
Zui low detectable concentration 0.002mg/L (enriched 100 times the oil concentration in water)
Detection limit 3SD <0.2 mg/L (solution concentration)
Baseline drift 1%/4h
Measurement errors of different ratios <5%
External dimensions 540×246×160(mm)
weight 13.5kg
power supply 220V±40V 50Hz±1Hz 40W
humidity <80%
temperature 5-35℃