brand | Self operated brand | application area | Environmental protection, chemical industry, biotechnology, petroleum, energy |
---|
WKL-3000B typeMicrocoulometric chlorine analyzerparameter
Technical Parameter |
Bias voltage range:0 ~ 500mv |
Standards |
ASTM D5808-18 Test method for detecting organic chlorine in aromatic hydrocarbons and related compounds using microcoulometry ASTM D5808-09a Standard Test Method for Chlorine in Aromatic Hydrocarbons and Related Chemicals by Microcoulometric Analysis ASTM D5194-18 Test method for trace chloride in liquid aromatic hydrocarbons SH/T 1757-2006 Determination of Organic Chlorine in Industrial Aromatic Hydrocarbons Microculometric Method GB/T 18612-2011 | Determination of Organic Chlorine Content in Crude Oil ASTM D4929-201Test Method for Determining Organic Chlorine Content in Crude Oil GB/T 23971-2009 Appendix BIn crude oilTotal chlorine contentdetermine Electric quantity method SH/T 0677-1999 Appendix A Determination of Total Chlorine Content in Crude Oil (Electricity Method) UOP779-08 Chloride in Petroleum Distillates by Microcoulometry YB/T 5022-2016 Crude Benzene Sulfur Chlorine Detection SY/T 7329-2016 Method for determination of organic chlorine content in oilfield chemicals |
Measurement accuracy:
Concentration value(ppm) |
Injection volume(μL) |
RSD(%) |
0.2~1 |
10 |
30 |
1~10 |
10 |
10 |
10~100 |
5 |
3 |
>100 |
5 |
2 |
Gas source requirements: ordinary nitrogen, ordinary oxygen
Power requirements:AC 220V±22V,50HZ±0.5HZPower:3.5KW
Temperature control range: Room temperature~1100Temperature control accuracy at ℃: ±1℃
External dimensions:700×480×540(mm)Weight:46Kg
WKL-3000BType microcoulombchlorineanalyzeradvantage:
Integrated structure, compact design, and beautiful appearance.
Controlled by computer, color LCD displays peak shapes.
Automatic data processing, report printing and output.
High precision PID regulation control(±1℃).
Overheating protection for cracking furnace, long service life(service lifeMore than 3 years).
The mixer is integrated with the main unit to prevent the quartz tube from getting stuck or broken.
The insulation and heat dissipation settings of the high-temperature furnace are effective.
Equipped with self checking functionAutomatically determine hardware faults or analyze system issues.
Cracking furnaceThe experimental temperature is set, controlled, and saved by the computer.
Adjusting bias voltage to achieve automation,As long as the target bias is set, unmanned and automatically adjusted.